Tx Planning Final Flashcards
T/F: Penumbra decreases w/increasing SSD
F
T/F: Penumbra increases with larger source size
T
Electrons can be used to treat tumors to a depth of _________cm deep.
5
T/F: Safety is no accident is published by ASTRO and is a recommendations for radiation therapy departments
T
Treatment planning includes the following steps: (select all that are true)
patient positioning
tumor delineation
beam arrangement
computing dose delivered and eval of treatment plan
transfer of the treatment planning information to delivery sysetm
All
T/F: If you tattoo a patient in SIM at the CT setup point but dosi moves isocenter to another location you will have to shift daily on the machine?
T
How do you verify your shifts from CT isocenter to treatment isocenter?
Subtract the coodinates in the X,Y, and Z direction from CT iso to plan iso
When determining parameters to optimize a patients treatment some of the parameters include: (select all that are true) Target volume Dose-limiting structures Treatment volume dose prescription dose fractionation dose distribution
All
T/F: OARs are contoured on a CT scan by the physician
F
T/F: IMRT is known as inverse planning
F
What modality gives the best spatial resolution?
CT
motion artifact can happen on scan from the following: (Select all that is correct) breathing motion cardiac motion bone motion table motion
breathing motion
cardiac motion
Match the tissue-type to the associated HU:
bone
water
air
bone = +1000 water = 0 air = -1000
At diagnostic energies the most common interaction in tissue is what? (select all that is correct)
photo electric
compton effect
pair production
photo electric
compton effect
Gross tumor + microscopic disease is called
CTV
T/F: Target volume is defined as the primary tumor + local spread and lymphatic spread
T
Commonly used imaging studies to help delineate the tumor is the following: (Select all that apply) CT PET SPECT MRI PMRI
CT
PET
SPECT
MRI
Irradiated volume is defined as the ____________ Isodose line
50%
What’s the energy range?
Co-60, 4MV, 6MV
Low-energy Megavoltage X-ray
What’s the energy range?
10-25MV
Medium or High energy Megavoltage X-ray
T/F: To treat with IMRT you must have MLC or compensator option on your treatment unit
T
T/F: Using low energy megavoltage x-ray beams while treating parallel-opposed beams, the body thickness should not exceed approximately 15cm?
T
Delivering diff dose to diff volumes in 1 course of tx over 33 fxs.
Dose painting
Select below the MLC size that would give you smaller beamlet options
1cm
0.5cm
0.5cm
Constraint or objective?
Conditions that are to be satisfied in any case
constraint
Constraint or objective?
Functions that measure the quality of a treatment plan
objective
Serial or parallel organ?
spinal cord
serial
Serial or parallel organ?
lung
parallel
T/F: Dynamic or sliding window IMRT delivery technique is best described as when the beam turns on and delivers dose, then turns off while the MLCs move to the next position then the beam turns back on. This pattern continues until all MU are delivered.
F
Yaw, pitch, and roll are what type of shifts?
Rotational
The interaction that happens with PET scans is what - Annihilation Radiation or Compton?
Annihilation Radiation
An ITV is created when there is movement of the tumor that needs to be accounted for.
T