Troubleshooting Flashcards

1
Q

How do you troubleshoot the coil?

A
  • Remote check
  • Swap coil (with backup)
  • Signal check tool (should light up if transmitting RF signal; if not lit up, still could be coil/cable/processor)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How do you troubleshoot the cable?

A

-More likely the issue instead of the coil

  • Remote check
  • Swap cable
  • Signal check tool (want to move it around to see if light starts flickering; may indicate frayed wire, etc.)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How do you troubleshoot the microphones?

A
  • Listening check
  • Replace mic covers
  • Swap processors
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How do you troubleshoot the processor?

A
  • Remote check
  • Swap processor
  • Connect to computer (if connects, try re-downloading MAPS or reprogram)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How do you troubleshoot the battery?

A
  • Check with remote
  • Clean contacts
  • Switch to disposable
  • Power optimization
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How do you troubleshoot the remote/fine tuner?

A
  • Verify pairing
  • Remote check (re-verify that the processor is trying to communicate with the remote)
  • Software configuration (was the patient given access to ___ parameter?)
  • Reset
  • Clinic remote
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

PATIENT COMPLAINT: physical discomfort

A

SOLUTION:

  • Coil: loose/replace magnet, spacer
  • Earhook: size, shape
  • Processor/battery: “hot spot” (get moleskin), referred pain
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

PATIENT COMPLAINT: retention

A
  • Size of the earhook
  • Magnet strength (headband use)
  • SnugFit
  • Earmold
  • Toupee tape
  • Patient’s hair
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

PATIENT COMPLAINT: intermittent function

A
  • Check skin flap thickness (headband, magnet, spacer, revision procedure)
  • Coil (movement, transmission)
  • Cable
  • Processor
  • Battery
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

PATIENT COMPLAINT: static

A
  • Coil
  • Cable
  • Microphones (listening check, mic covers)
  • Processor/MAP (telecoil?, M/C levels too high?)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

PATIENT COMPLAINT: very soft volume

A
  • Skin thickness at coil
  • Mic covers
  • Microphones
  • Batteries (end of life?)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

PATIENT COMPLAINT: manual adjustments don’t work

A
  • Remote/fine tuner paid?
  • Processor configuration in software?
  • Telecoil activated?
  • Processor buttons locked?
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

PATIENT COMPLAINT: cracked/chipped case

A

-Order replacement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

PATIENT COMPLAINT: no sound

A
  • Batteries
  • Processor on?
  • Microphones
  • Coil/cable
  • Back-up device
  • Internal device? (connected through computer? detecting internal implant?, try clinic processor)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly