Troubleshooting Flashcards
1
Q
How do you troubleshoot the coil?
A
- Remote check
- Swap coil (with backup)
- Signal check tool (should light up if transmitting RF signal; if not lit up, still could be coil/cable/processor)
2
Q
How do you troubleshoot the cable?
A
-More likely the issue instead of the coil
- Remote check
- Swap cable
- Signal check tool (want to move it around to see if light starts flickering; may indicate frayed wire, etc.)
3
Q
How do you troubleshoot the microphones?
A
- Listening check
- Replace mic covers
- Swap processors
4
Q
How do you troubleshoot the processor?
A
- Remote check
- Swap processor
- Connect to computer (if connects, try re-downloading MAPS or reprogram)
5
Q
How do you troubleshoot the battery?
A
- Check with remote
- Clean contacts
- Switch to disposable
- Power optimization
6
Q
How do you troubleshoot the remote/fine tuner?
A
- Verify pairing
- Remote check (re-verify that the processor is trying to communicate with the remote)
- Software configuration (was the patient given access to ___ parameter?)
- Reset
- Clinic remote
7
Q
PATIENT COMPLAINT: physical discomfort
A
SOLUTION:
- Coil: loose/replace magnet, spacer
- Earhook: size, shape
- Processor/battery: “hot spot” (get moleskin), referred pain
8
Q
PATIENT COMPLAINT: retention
A
- Size of the earhook
- Magnet strength (headband use)
- SnugFit
- Earmold
- Toupee tape
- Patient’s hair
9
Q
PATIENT COMPLAINT: intermittent function
A
- Check skin flap thickness (headband, magnet, spacer, revision procedure)
- Coil (movement, transmission)
- Cable
- Processor
- Battery
10
Q
PATIENT COMPLAINT: static
A
- Coil
- Cable
- Microphones (listening check, mic covers)
- Processor/MAP (telecoil?, M/C levels too high?)
11
Q
PATIENT COMPLAINT: very soft volume
A
- Skin thickness at coil
- Mic covers
- Microphones
- Batteries (end of life?)
12
Q
PATIENT COMPLAINT: manual adjustments don’t work
A
- Remote/fine tuner paid?
- Processor configuration in software?
- Telecoil activated?
- Processor buttons locked?
13
Q
PATIENT COMPLAINT: cracked/chipped case
A
-Order replacement
14
Q
PATIENT COMPLAINT: no sound
A
- Batteries
- Processor on?
- Microphones
- Coil/cable
- Back-up device
- Internal device? (connected through computer? detecting internal implant?, try clinic processor)