Topic 4 Flashcards

1
Q

Normal Phase HPLC

A

Non-polar mobile phase
Polar stationary phase
Least polar elutes first

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Reverse phase HPLC

A

Polar mobile phase
Non-polar stationary phase
Most polar elutes first

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Ion-exchange column

A

Separation of ions and polar molecules

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Size-exclusion column

A

Separation of molecular size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

UV-Vis

A

Compound must be UV-sensitive

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Advantages of UV-Vis

A

1) Inexpensive
2) Applicable to many organic compounds
3) Robust

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Advantages of Evaporating Light Scattering detector

A

1) Applicable to all analyte species
2) Suitable for non UV sensitive compounds
3) Suitable for heat-sensitive compounds

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Advantages of Mass Spectrometer

A

1) Extremely sensitive
2) Able to obtain MS profile

Limitation - only for ionisable compound

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Abnormal pump pressure (High)

A

Caused by:
1) Column clogging
2) Precipitation of solvent lines

Solutions:
1) Flush with strong solvent
2) Flush with solvent the precipitate is soluble in

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Abnormal pump pressure (Low)

A

Caused by:
1) Leaks
2) Air bubbles between pump and mobile phase reservoir

Solution:
1) Check for leaks
2) Purge solvent lines

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Shifts in retention time

A

Caused by:
1) Column aging
2) Contamination of mobile phase
3) Insufficient equilibrium time

Solution:
1) Replace column
2) Flush column
3) Longer column equilibrium times

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Detection issues

A

Noisy peaks
Drifting baseline

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Noisy Peaks

A

Caused by:
1) Uv lamp reaching end of life
2) Air bubbles
3) Intermittent pulses from pump

Solution:
1) Change uv lamp
2) Flush lines / check degasser functioning
3) Check pump fuctioning

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Drifting baseline

A

Caused by:
1) Column bleeding
2) Mobile phase contamination

Solutions:
1) Replace column
2) Reprepare mobile phase

How well did you know this?
1
Not at all
2
3
4
5
Perfectly