Tonometry Flashcards
What is the number of a target illuminator replacement bulb?
A. 12419
B. 19241
C. 11249
D. 14129
A. 12419
What drug is used during tonometry to anesthetize the eye?
A. Proparacaine
B. Paredrine
C. Phenylephrine
D. Tropicamide
A. Proparacaine
Killing of the HIV and other viruses on the applanation prism is BEST accomplished by soaking in
A. Diluted Betadine
B. Gentamycin solution
C. Isopropyl alcohol
D. Sodium hypochlorite (chlorine bleach)
D. Sodium hypochlorite (chlorine bleach)
Which type of tonermeter is mounted on a slit lamp?
A. Applanation
B. Indentation
C. Perkins
D. Air puff
A. Applanation
The calibration of the Goldmann applanation tonometer should be checked at which settings?
A. 0,1,2
B. 0,2,4
C. 0,2,6
D. 0,4,8
C. 0,2,6
What is the diameter of the plastic prism comprised of a flat surface within the applanation tonometer?
A. 1.5 MM
B. 3.0 MM
C. 6.0 MM
D. 7.0 MM
B. 3.0 MM
In applanation tonometry, excessive width of the mires can indicate
A. Improper alignment of the prism
B. A need for recalibration of the instrument
C. Too little fluoresceine
D. Too much fluoresceine
D. Too much fluoresceine
What is the MAIN advantage of the applanation tonometer compared with the indentation tonometer?
A. Applanation is a more modern technique
B. Applanation tonometry causes less patient discomfort
C. It is more accurate
D. The intraocular pressure change is larger as compared to indentation
C. It is more accurate