The Ventilator Flashcards
What is inspiratory airflow?
Occurs as a result of a pressure gradient where the extrapulmonary pressure is greater than the intrapulmonary pressure
What is expiratory airflow?
Occurs as a result of a pressure gradient in which intrapulmonary pressure is raised to above extrapulmonary pressure
What is tidal volume?
- It is the amount of air entering or inhaled or leaving by exhalation with each breath
- When a ventilator mode uses a target tidal volume in adults the expected settings should be between 6-8 ml/kg of the patient’s ideal weight. The goal is not to exceed the plateau pressure of 30 cm of H2) to prevent barotrauma.
- The most common practice is to use 50 ml increments and approximate the tidal volume and adjust to effect. Thus tidal volumes can be 400, 450, 500 ml and so on.
What is plateau pressure?
Plateau pressure is the pressure applied to the alveoli and small airways. Lung barotrauma can occur when there is over-inflation causing damage to tissues and circulation. Air embolism can occur.
What is a “rate” on the ventilator?
The rate is the number of breaths delivered each minute
- A mechanical breath cycle can be defined by either the tidal volume amount or target OR by a pressure target
- Rates are adjusted according to blood gas analysis to achieve pH and PaCO2 goals
What is FIO2?
Fraction of inspired oxygen (FIO2) is the decimal value produced by dividing the partial pressure of oxygen (PO2) by the total pressure of the mixture.
Therefore:
- FIO2 of room air is approximately .21. Using a decimal value rather than a percentage is the correct expression.
- FIO2 of 100% is 1
- In the hospital setting however, the most accepted way to express oxygen concentration that you see will likely be %
What are 2 types of inspiratory cycles?
- Volume cycled: pressure limited (volume target). Preset tidal volume is delivered unless a set pressure limit is reached. At that point, the cycle is terminated
- Pressure cycled: volume limited (pressure targeted). A preset pressure is delivered unless a volume limit is reached terminating the cycle.
What is the minute volume/ventilation?
It is the total volume of new air that enters the respiratory passages per minute
What are the benefits of a high PEEP?
It prevents the return of intrapulmonary pressure to equal extra pulmonary pressure at the end of expiration. It accomplishes an incomplete expiration leaving a higher amount of air in the lungs or functional residual volume (FRV)
What are the effects of an increase in functional residual volume (FRV)?
- Better gas exchange as more alveoli are left open throughout the ventilatory cycle
- There are additional alveoli that will open at peak inspiration because the delivered tidal volume is “stacked” on the FRV.
- A higher PaO2 may be achieved without having to increase the FIO2
- Typical PEEP settings are 5-10 cm H20 although higher settings may be needed for patients that have low compliance conditions such as acute respiratory distress syndrome
- Risks of PEEP include barotrauma, impedence of central venous return that can result in a lower cardiac output and increased intracranial pressure
What is CPAP and when is it used?
- Same as PEEP but with no mechanically delivered inspirations
- Patient must be able to breathe without assistance
- Has the same risks and benefits as PEEP
What causes a high-pressure ventilator alarm?
A high-pressure alarm will occur when the proximal airway pressures exceed the limits that are set. Alarm conditions can occur when there is:
- secretion accumulation
- the patient is coughing
- there is spontaneous dysynchrony
- decreasing lung compliance
- pneumothorax
- airway occlusion
What causes a low-pressure ventilator alarm?
A low-pressure alarm occurs when the proximal airway pressure does not reflect the current ventilator function. This can happen with disconnected tubing and/or ETT leaks
When is a ventilator “trigger”?
Determines how the ventilator initiates a breath. Can either be pressure triggered or volume triggered
What is “cycling” when using a ventilator?
Tells the ventilator when to terminate the breath to allow expiration