Test 2 Flashcards
The recognizability factors in an image depend upon the ____ integrity of the projection
Geometrical
Why is penumbra present to some degree in all practical images?
Because no source of radiation is an infinitely small point, (all consist of areas).
The umbra of very small details in an image can completely disappear by what geometrical process
If the focal spot is significantly larger than the object being projected.
misalignment of the x-ray beam, part and image receptor results in what undesirable image quality
Shape distortion
Given a 2mm focal spot size and SOD of 90 cm and on OID of 30 cm, what is the geometrical unsharpness of the image
0.67 mm
Given an 0.5mm focal spot and SID of 150 cm and an OID of 15 cm what is the geometrical unsharpness of the image
0.055 mm
what is the relative sharpness for an SOD of 90 cm and an OID of 15 cm
6
an original exposure used an SOD of 120 cm and OID of 12 cm. IF the SOD is increased to 150 cm and the OID is reduced to 10 cm by what factor will the sharpness be improved
By 1½ times or 50%
With an SID of 60 cm and an OID of 40 m by what factor will the image be magnified
3 X
an image measures 16cm across. A 60 cm SID was used with an SOD of 40 cm. What is the size of the original object
3 X
what is the percentage magnification for an image with a magnification factor of 1.44
44%
The radiographic image of a particular bone turns out 4 times longer than the real bone, but only 2 times wider. what is the factor of magnification? what is the factor of shape distortion
Magnification = 2 X; Shape distortion = 2 X
The resolution of line pairs from a test template can be reduced by either of which two processes
Blurring or loss of contrast
The ability to distinguish any two adjacent details as separate and distinct details is the definition for what aspect of an image
Resolution
What is the typical range of anode bevel angles fro a general purpose x-ray tube
15-17 degrees
rather than using the line-focus principle, why don’t we just focus down the beam of electrons so that the actual focal spot is 0.5 mm
Because this would damage the anode by concentrating the heat load
the projected focal spot is largest at the end of the ___ image receptor
Cathode
explain why the anode heel effect occurs
Because the anode material itself acts as a filter for x-rays
x-ray tubes are conventionally installed with the anode to your ____ as you approach the x-ray table
left
for a radiograph of the AP foot or femur, it is best to position the patient on the x-ray table with his head to your _____
right
at a shorter SID, the collimator must be opened up to cover the same field size, so the anode heel effect ____
increases
what is the most compelling reason to consider focal spot size as the controlling factor for sharpness in the remnant beam images
The focal spot affects only image sharpness, it does not affect any other image qualities
why is the small focal spot not available when high mA stations are engaged
It is locked out from these high mA stations to prevent overloading the anode with concentrated heat
the actual focal spot is best measured along the
Beveled anode surface
explain how a focal spot which is much larger than an anatomical detail can cause it to disappear from the resolved image
When the focal spot is substantially larger than an image detail, the penumbra can spread so much inward that the umbra portion of the image is completely lost.
If the size of the focal spot is cut exactly in half, the sharpness of the image carried by the remnant x-ray beam will change which way and by precisely how much
Sharpness will increase to precisely double the original
where within the penumbra does the human eye locate the edge of an image detail
At the center of the penumbra
geometrical penumbra is caused by the ____ absorption of xrays as a function of the total absorption which the anatomical structure is capable of
Partial
decreasing the SID _____ penumbra
Increases
list four benefits of increased SID
1) Patient dose is slightly reduced, 2) Spatial resolution (sharpness) is improved,3) Magnification of anatomy is reduced, 4) More anatomy can be included within the field of view
if a change in the SID has resulted in the exposure increasing by 4 times, all other factors equal, the SID must have been changed by what amount?
It must have been cut to ½ the original
a technique chart lists 0.05 second for a procedure using 100 cm SID at a new SID of 200 cm, what new time would be required to maintain the exposure at the image receptor
0.2 seconds
for any change greater than ____ percent in the SID, radiographic technique factors should be adjusted to compensate and maintain exposure
15%
when changing from 180 to 100 cm, change the mAs to ____ the original
1/3 (0.31)
image sharpness is directly proportional to which radiographic distance
SOD (source-to-object distance)
in the interest of image sharpness, generally the ____ Feasible SID should be used
Maximum
as a rule-of-thumb, when changing from a 100cm SID to a 125 cm SID, increase the mAs by ______
One step
an SID of 128 cm results in an exposure at the receptor plate of 10 mGy, Using the inverse square law if the SID is changed to 92 cm, what will the new exposure be
19.4 mGy
at an SID of 244cm, 120 ma results in a satisfactory exposure. if the SID is changed to 75 cm what new mA must be sued to maintain the exposure at the receptor plate
11.4 mA (approximately 10 mA)
at an SID of 200 cm, 25 mAs results in a satisfactory exposure. If the SID is changed to 90 cm , what new mAs must be used to maintain the exposure at the receptor plate
5.06 mAs (approximately 5 mAs)
in general radiography, the ____ possible OID should always be used
Minimum
The greater the OID the ___ the subject contrast in the remnant beam image
Greater
The greater the OID the ____ the sharpness of detail in the image
Less
at greater OID, ____ radiation is allowed to spread out more, while ____ radiation remains at the same concentration
Secondary or scatter; primary
magnification is directly proportional to the ____ ratio
SID/SOD
image sharpness is directly proportional to the ____ ratio
SOD/OID
If the SID is 100cm and the SOD is 50cm , the projected image of the object will be magnified by a factor of _____
2
an original technique uses an SID of 100 cm and an OID of 5cm. If the OID were increased to 7.5cm what new SID would be required in order to completely eliminate the blurring effects of the OID change
150 cm
an original magnification ration is 50 percent. if the SID and OID are both tripled, the new magnification ration will be _____
50% (the same)
if both SID and OID are increased by 15 cm, what will be the net effect on the exposure intensity at the image receptor plate
Decrease
If both SID and OID are increased by 20 cm, what will be the net effect on the subject contrast in the image at the receptor plate
Increase
why is off centering of the CR identical in its effects to angling the x-ray beam
Because it places objects into diverging beams which are at an angle to the CR
why do positioning atlases recommend a 15-degree angle for the AP projection of the sacrum, when the sacrum actually lies at 30-35 degree angle
To minimize shape distortion (and to desuperimpose the symphysis pubis)
which of the following would be most distorted by a 30 degree CR angle: the head of the femur, the shaft of the femur, or the sternum
The head of the femur
if a flat object is tilted in relation to the receptor, and the CR is angle perpendicular to the object, what type of distortion if any occurs
Elongation
the distorting effects of off centering are indirectly worsened when the SID is _____
Decreased
If the R is angled 20 degrees, the TTD should be changed from 100 cm to ___ in order to maintain the exposure level
90 cm
List the four aspects of beam part receptor alignment which impact upon shape distortion
1) Shape and size of the object, 2) centering of the object relative to the x-ray beam, 3) angle formed between the object and the x-ray beam, 4) angle formed between the object and the image receptor plate
why does the cranium distort more than the head of the femur with an angled projection
Because of the cranium’s much greater thickness
what is the angle of x-ray beam divergence at a point 10 m cephalic to the CR
7 degrees (cephalic)
for a lateral projection of the lumbar spine, why is it not desirable to build the spine up all the way to a horizontal position
Normal divergence of the x-ray beam lines up with the intervertebral joint spaces when there is slight sag in the spine
what are the four geometrical objectives of radiographic positioning
1) Desuperimpose obstructing anatomy, 2) minimize magnification, 3) minimize distortion, 4) maximize sharpness
why are multiple projections at different angles necessary for fracture radiogrpahy
To fully demonstrate alignment of bones
contrasty anatomy that obscures the anatomy of interest is a form of image ____
Noise
penumbra in the image is ____ by motion of the par, x-ray tube or recptor
Increased (spread)
how might eh x-ray tube accidentally be moving during an exposure
If it is not locked in place
by superimposing various image densities, severe motion can destroy image ___ as well as sharpness
Contrast
streaks or circular artifacts caused by tomographic movement are classified as ____ images
False
list the three methods of minimizing motion during radiographic exposures
1) Minimal exposure time, 2) securing patient cooperation, 3) use of mechanical restraining devices
to freeze motion, exposure time should not exceed ___ second
0.033 (1/30) second
What are the three components of recognizability
sharpness, magnification, shape distortion
what does SID stand for
source to image distance
what does SOD stand for
source to object distance
what does OID stand for
object to image distance
another name for sharpness of detail
spatial resolution
area on the anode where the electrons hit
actual focal spot size
if the angle of the bevel increases the focal spot ____
increases
the clearly defined portion of a projected image
umbra
blur along the edges of an image
penumbra
formula to calculate for penumbra
penumbra = Fs x OID/ SOD
Smaller focal spot sizes produce ____ penumbra which in turn produces better ____ _____
less, spatial resolution
Why do we not use a small focal spot size for all images
it burns out the anode due to too much heat
If we increase SID what does it do to spatial resolution
increases
if we increase SID what does it do to magnification
decreases
what is the formula for magnification factor
image size/ object size or SID/SOD
what is the formula for object size
image size (SOD/SID)
what two types of distortion are there
magnification and shape