Systems Of Dose Calculations Flashcards
Linacs are calibrated so that 1 MU delivers 1cGy for a standard reference __ at a standard reference __ at a standard source-to-calibration point
field size, depth
MU reference depth used in most clinics followed by the alternative reference depth
Dmax is most common; 10cm depth is the alternative
MU = __ at a point ÷ __ at that point
Dose, dose rate
What things affect (impact) dose rate?
Output (distance specific)
Attenuation factors
Tissue attenuation (PDD, TAR, TMR, TPR)
Beam attenuation (tray factor, wedge factor)
What special eqn to use to find dose to a point using PDD
Given Dose = Tumor Dose ÷ (PDDsubTD × 100)
What is the ratio method eqn to find dose to a point uaing PDD?
DosesubA/PDDsubA = DosesubB/PDDsubB
What is the eqn to calc MU setting using PDD?
MU = Rx / everything else (multiplied)
PDD is used for SSD or SAD?
SSD
FIXWhat eqn to use to find dose to a point with TAR, TMR, and TPR?
DosesubA = (DosesubB/TARsubB)(what is scpd??????
TAR, TMR, TPR is used for SSD or SAD?
SAD
Rx/PDD = ?
Dmax
Define given dose
Dmax
Define applied dose
Dmax
Reference dose rate calibration parameter?
1 cGy to Dmax at 100cm SSD = 1.0 cGy per MU
When do you use Mayneord Factor in calculations?
When you treat at extended distances. Must use Myaneord F to correct PDD (non inverse square)