Structure Flashcards
What is the ESU ?
Energy storage unit, only fitted on 25KW unit. 2 units, left and right. Allows 30s fluoro
What is the generator called ?
Polydoros M25
Max tube current for single images with 25KW generator
250mA
Focal spot sizes
0.3/0.5
Left digit on D80 board displaying ?
Intermediate voltage
How do you discharge ESU or ELKO buffers ?
D80.S1 position 2. Upper position
What is the MCU ext used for ?
Reading end switched/brakes of angular, orbital and horizontal drives
What operating voltage is required for the FD ?
+24V
What max value is allowed for the protective ground check ?
0.3 ohms
What is indicated by the red LED on the detector ?
Communication to the CoCa board is down.
Which filter can be used for detector calibration ?
0.6mm Cu, delivered with unit
Where is the mains adaption carried out ?
By jumper settings on the D56
What is the service switch to enable radiation ?
D80.S3, position 1 = service position
What is the function of the right hand display on D80 ?
Generator state
What is an EVAR ?
Endovascular Aortic repair
What does AAA mean ?
Abdominal Aortic Anurism
What is the customer login under service and why do they need a login ?
- admin2
- To calibrate the detector, or change password
Which input fields of a 30x30cm detector are selectable ?
- 30x30, 20x20 and 15x15
What WLAN standards are selectable ?
IEEE 801.11 a,b,g,h
What is the matrix size of a 30x30cm flat panel ?
1536x1536
What frequencies are used to control the Inverter ?
18-50khz
What is the SID of the anti-scatter grid ?
115cm
What type of detector calibration is carried out on site ?
Service
What is the attenuation of the UNFORS dose meter for DEXI calibration ?
1.62