sterilisers Flashcards

1
Q

what is the guidance document?

A

SHTM 2010

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

daily info for sterilisation

A
date
time started
operator
cycle no
sterilisation temp
daily tests performed
sterilisation successful
machine identified cycle completed successfully
Type B - Bowie Dick successful
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

who carries out the daily and weekly tests?

A

on site staff

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

temp

A

134-137 degrees

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

time

A

min hold time 3mins

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

pressure

A

pressure equivalent gauge (from atmospheric) 2.05 to 2.35 bar
pressure equivalent absolute (from absolute 0) 3.05 to 3.35 bar

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

do all sterilisers achieve the same temp and pressure?

A

yes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

daily tests

A

wipe door seal with clean, disposable, damp, non-linting cloth and carry out any other checks required by the manufacturer, check seal for damage/contamination

check load carrier and chamber for damage/debris/clean

fill and drain water reservoir

when switching power on, check that the ventilation louvers are not covered to avoid overheating

if recommended by the steriliser manufacturer, preheat the steriliser chamber before performing daily tests

record completion of the daily checks in the steriliser logbook

Type B: Bowie Dick (steam penetration) test

record daily sterilisation temps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

which extra daily test is required for type B sterilisers?

A

Bowie Dick (steam penetration) test

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

sterilisation

A

a process undertaken to render a steriliser load sterile

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

sterile

A

condition of a medical device that is free from viable MOs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

weekly tests

A

Type B - air detector function test and air leakage test
ACT
daily tests
weekly safety checks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

weekly safety checks

A

examine door seal for signs of wear or damage

examine security and performance of the door safety features inc the hinges and locking mechanism as detailed in the MIs

if a fault is detected in the door seal or safety features, ensure this is corrected before carrying out weekly tests or using the steriliser

record satisfactory completion of the weekly safety checks in the steriliser logbook

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

type N steriliser - how does it work?

A

‘gravity displacement’
air passively displaced by steam
purified water introduced to the chamber and heated
as it boils and turns to steam, the chamber pressure increases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

what instruments can be processed in type N?

A

rigid steel solid instruments

no wrapped/channeled/lumened

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

stages in type N

A

air removal - heating time
sterilising - holding time
condensing - cooling time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

what instruments can be processed in a type B?

A

wrapped and non-wrapped (make sure not to damage wrappings)
channelled and porous instruments
porous loads e.g. cotton swabs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

loading recommendations

A

no overlapping/shadowing
do not overload machine
assemblies should be disassembled beforehand
utilise the available basket/load carrier

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

type B - how does it work?

A

porous load/vacuum capable
has active air removal creating a vacuum in the chamber in its initial stage
so steam can contact all surfaces of device
because there is no air in the chamber, the vacuum rapidly draws the steam in from the generator
after sterilisation the machine draws another vacuum for drying, allowing water to vapourise at lower temps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

B type - after processing

A

wrapping intact - ‘sterile’ at point of use

can store in env controlled atmosphere that is free from contaminants and moisture

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

N type - after processing

A

‘sterilised’
wrap as soon as remove from steriliser and transport to procedure room
either for immediate use or for non-sterile storage, transport and application

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

why should instruments be left to cool down after first?

A

if put hot instrumentation into transport box might cause condensation to build up, potentially damaging wrappings

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

for sterilisation of wrapped goods what must the residual gas not exceed?

A

3.5 ml per 100ml of condensed steam

air detector incorporated as a safety device to monitor the adequacy of the air removal process

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

what is residual gas?

A

remaining air and gases carried in the steam after the air removal stage

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
type S
products specified by the manufacturer of the steriliser inc non-wrapped solid products and ≥1 of the following: porous products, small porous items, hollow load products type A, hollow load products B, single wrapped products, multiple layer wrapped products vacuum pump/some other active method to remove air from the chamber and load
26
what should dental handpieces be sterilised using?
a vacuum process | type S
27
obligations under H+S and Consumer Protection Act
ensure equipment: - complies with safety requirements - installed and maintained appropriately - validated and routinely tested - operated only in accordance with MIs and by properly trained operators
28
water and steam quality
good quality water to make good quality steam | water low in minerals to prevent depositing minerals onto the instruments (limescale) and making them harder to clean
29
waters recommended by manufacturers for use in sterilisers
``` = pure water Reverse osmosis distilled sterile deionised ```
30
basic stages of operation of Type B
``` 1 - air removal (negative pulsing) 2 - steam admission (positive pulsing) 3 - equilibration 4 - sterilisation 5 - drying 6 - air admission (0.2micron fitted inlet) ```
31
type B : 1 - heating up of water and air removal
water enters an external boiler known as a steam generator to a pre-determined level from the clean tank heating element switched on (generator heating water) machine is pulling a pre-determined vacuum (neg pressure) as water reaches bp, steam generated. Element stays on until the steam reaches sterilising temp once vacuum achieved to a pre-determined level the steam generator injects steam - 3 times (neg pulsing) on 3rd injection of steam vacuum cuts off and the steam is allowed to build up in stages (positive pulsing) to the sterilising temp and pressure
32
type B: 2 - sterilising
once all air removed and temp/pressure achieved it is held for a holding time (min 3mins) sterilising temp controlled by a combination of pressure transducers and sump and chamber temp sensors at end a valve opens to allow the condensate steam to enter the dirty tank
33
Type B: 3 - drying
B unlike N has a proper drying stage pulling a vacuum whilst keeping the chamber hot - remove steam some sterilisers do vacuum pulses at this stage whilst others only pull the one deep vacuum negative pressure
34
Type B: 4 - cooling (air admission)
after a predetermined time the vacuum is switched off and the chamber allowed to cool when the chamber reaches atmospheric pressure and a set temp (typically 80 degrees) the door is unlocked and can be opened
35
mass
amount of matter an object contains (kg)
36
weight
force caused by gravity on an object (N) | - weight of a mass 1kg approx 10N
37
pressure
force acting normally per unit area | 1Pa = 1N/m2
38
atmospheric pressure
often given in millibars where standard sea level pressure defined as 1000mbar = 1 bar when we speak about vacuum for sterilisation normally use mbars when we speak about positive pressure we use bar e.g. 2.3 bar
39
automatic controller
device within the steriliser that controls the sterilisation cycle to ensure it is working - ACT
40
how often should ACT be done?
daily: either using the sterilisation cycle parameter values recorded on the printout or electronic data logger, or by manually observing and recording the cycle parameters if there is not a suitable recorder fitted weekly: manual ACT
41
can the ACT be done when sterilising a standard load?
yes unless also carrying out a steam penetration test for a vacuum steriliser at the same time - usually 1st cycle of day
42
when must a daily manual ACT be carried out?
if a suitable recorder is not fitted as normally would just be done daily with steriliser recorder - a printer or electronic data logger
43
ACT using a recorder
run a sterilisation cycle with a standard load or an empty chamber (the chamber must be empty if also carrying out a steam penetration test in a vacuum steriliser) at the end of the cycle, check printout/data logger to ensure that the recorded cycle parameters (temp, pressure, hold time) are within the specified range for the cycle and comparable to the values obtained at validation - keep a record of the values if ACT unsatisfactory (i.e. recorded temp, pressure or hold time are not within the specified range for the cycle), record the test as a fail and do not use until fault resolved - return any instruments that were loaded in the steriliser to the start of the decon cycle sign logbook
44
how often should the Bowie-Dick (steam penetration test) be carried out?
daily
45
which type of steriliser needs Bowie-Dick (steam penetration test) ?
type B
46
steam penetration test types
Bowie and Dick test pack - chemical indicators | Helix/hollow load test
47
Bowie and Dick test pack
chemical indicators test sheet fitted between hundreds of sheets of paper yellow - dark blue if air removal part of cycle not performing correctly - yellow spotting on the sheet pink coloured circle on outside changes colour once processed - purple
48
how to carry out Bowie-Dick (steam penetration test)
``` place on bottom shelf of load chamber start cycle cycle much shorter and doesn't have same drying time remove pack open, find test sheet identify colour change ```
49
Helix/hollow load test
tests for steam penetration by letting us know that the steam was able to contact the chemical indicator by travelling through the entire length of tubing attached to the device
50
is a strip inserted into the pack when processing instruments acceptable?
no - not considered to show us anything except a pack of instruments has been through a sterilisation cycle - not a steam penetration test
51
main fct of steam penetration test
test steam penetration, also informs temp achieved, adequate air removal to show steam will rapidly and evenly penetrate a test device that is similar to the intended load
52
where can you get information on recommended steam penetration test device and indicator?
MIs
53
carrying out a steam penetration test
start of day - select usual cycle and perform test with chamber empty apart from the test device, following test device MIs record pass/fail in steriliser logbook if fail - repeat. A 2nd unsatisfactory test result confirms fault - arrange for a MP (sterilisers) to investigate and don't use until resolved
54
which steriliser needs an air leakage test?
type B
55
how often should the air leakage test be carried out?
weekly
56
other names for air leakage test
vacuum leak rate chamber integrity test vacuum test
57
why is it important air doesn't leak into the chamber in a type B?
if air leaks into chamber at a higher rate than specified by manufacturer, it might interfere with steam penetration into load, and as air will not have passed through the bacteria retentive filter, risk of recontamination of load air cannot be heated the same way as steam and therefore cannot achieve the required temp, or easily maintain that temp either
58
purpose of the air leakage test
verify that: during negative pressure stage it removes all air within the chamber at the start of the cycle when reach low pressure set point and vacuum pump stops operating - machine continues to maintain same pressure and no air is able to enter the chamber from any part of the machine = verifies that all valves and connections made to the chamber are sealing properly, the vacuum pump is capable of reaching the low set-point that we require and that the door is also sealing onto the chamber correctly
59
carrying out an air leakage test
chamber must be cold, dry, empty close door, machine locks it and cycle starts machine approaches set point pressure - then stabilises for 5mins(potential for pressure fluctuations), then measures any differential in the pressure over the following 10mins - max permitted change is 13mbar over 10mins or 1.3mbar per min. Max differential <2% once 10min over - see pressure start to increase as valve opens to allow filtered air back into the chamber - happens at beginning and end of cycle - ensure air can't be drawn in while cycle running (contamination) - check machine cycle data - displays all relevant values record results in steriliser logbook
60
automatic air leakage test
if an automatic test is available, do once a week (FMIs), generally with a cold chamber preferable to have a steriliser that is capable of performing an automatic test as otherwise a test person(decon) is required to perform a weekly manual test note that some manufacturers specify that an air leakage test is carried out each day before steam penetration test
61
why does the temp remain low in an air leakage test?
temp remains low as not a stage that has steam in it
62
how does machine display pressure in an air leakage test?
- pressure absolute displays atmospheric pressure as 1 bar (1000mbar) - machine displays it as pressure gauge, showing atmospheric pressure as 0 - removes air until reaches set point -0.87bar or -870mbar
63
what will any air leakage cause?
any air leakage will cause a change in the chamber pressure
64
if user cannot perform an ADFT, what needs to be done?
require a TP(D) to visit weekly to do it | - will add significantly to costs so check before purchase
65
which steriliser requires an ADFT?
B type
66
freq of an ADFT?
weekly
67
what machines are fitted with an air detector?
safety device | sterilisers that actively remove air from the load before sterilisation (e.g. using a vacuum pump)
68
what is an air detector?
generally a temp probe (/pressure transducer) in a difficult to reach part of the chamber, potentially at the bottom of a tube attached to the chamber positioned somewhere that will be difficult to remove air from - the resulting presence of air would mean that the temp at the probe would be lower than that of the temp the chamber probe would be measuring from the steam - temp differential checked and if it is outwith limits the cycle will fail
69
why is air in the chamber detrimental?
doesn't maintain temp the same way that steam does, and is also much harder to heat if air present during cycle could mean not every part of chamber and load being processed will reach required temp for required time - could allow MOs to survive
70
how much air is allowed to enter in ADFT?
small measured bleed of air is allowed to enter the chamber recommended rate 11mBar/min - v small - approx 0.5% of the pressure that we are working with - even v small amount of air has potential to affect cycle set points
71
after an ADFT
retrieve data from machine | record in steriliser logbook
72
which cycle can be used for the ACT?
1st production cycle
73
B type ACT things to record
date, time, cycle number, operator more data - cycle more complex record info for: vacuum stage at start, steam introduction and sterilisation stage and post-vacuum/drying stage. Time for each stage record readings at each peak, low pressure and high pressure - pulsing: record low pressure, high pressure and temp + pulsing: record low pressure, high pressure and temp sterilisation temp and pressures - at start of stage and then at 1min intervals until it finishes
74
N type ACT things to record
date, time, cycle number, operator record e.g. cycle times and sterilisation temps, times for diff stages record temp and pressure (if it is displayed) of the sterilisation stage at 1min intervals after create own ACT record can retrieve the machine cycle data as well and check both records against each other
75
vacuum steriliser - quarterly
``` weekly safety checks Bowie Dick test air leakage test ADFT calibration of test instruments ACTs and verification of calibration small load thermometric test calibration checks (post-testing) produce cycle graphs to make sure all cycle set points are being achieved ```
76
sterilisers - what must be done annually?
annual validation