Standards of a BVM Flashcards

1
Q

list the 12 standards of design for a BVM

A
  1. deliver at least 85% o2 with flow of 15LPM
  2. operate at extreme temps (-180 to 600dgC and RH of 40-96%)
  3. adult: must at least deliver Vt 600ml; Cs of 0.02 L/cmh2o and R of 20 cmh2o/L/sec
  4. neonates: must at least deliver Vt 200-300mls

pediatric: must at least deliver Vt 400-500mls

  1. flows 0-30LPM without jamming non-rebreathing valve
  2. malfunctioning valve d/t foreign obstruction must be restored w/in 20sec
  3. 15:22mm connector
  4. no pressure limiting system on adult BVM
  5. pressure release valve at 40 cmh2o (+/- 10) for paediatric BVM
  6. pressure release valve at 40 cmh2o (+/- 5) for neonatal BVM
  7. override capability & audible signal if pressure limiting/venting system in place
  8. operate properly after being dropped from height of 1m onto concrete