Which of the following occurs as the scan line density decreases?

A

Interpolation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Modern day ultrasound systems eliminate grating lobes in which of the following ways?

A

Electronically and mechanically

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Calculate the maximum frame rate if scanning to a depth of 10 cm, using a single focus and assuming there are 100 lines per frame.

A

77 frames per second

Frame rate formula = (77,000)/(LPF x Depth x No of foci)

LPF = lines per frame

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Which of the following is an assumption made by the ultrasound system?

  1. The strength of the returning echo indicates the organ echogenicity only
  2. Sound travels at a velocity related to compressibility and density
  3. Reflector distance is dependent on the proximal tissue
  4. Sound travels in a non linear fashion
A

The strength of the returning echo indicates the organ echogenicity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

This image displays areas of increased and decreased brightness as the result of several off axis reflections. Which of the following controls will improve this image?

A

Persistence

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Which of the following can produce a reverberation artifact?

  1. Bones
  2. Vessels
  3. Gas
  4. Stones
  5. Metal
A

All of these are correct

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

When scanning through ha lesion, the reflectors posterior to it appear closer to the probe than indicated on a comparable CT scan for the same patient. What does this tell the sonographer about the speed of sound through this lesion?

A

The speed of sound is faster than 1540 m/s

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Which artifact is demonstrated in this image?

A

Range ambiguity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the cause of the artifact seen in this image?

A

Focal banding

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Which of the doppler shifts listed is the minimum value that will result in aliasing with a PRF of 10 kHz?

  1. 3 kHz
  2. 5 kHz
  3. 6 kHz
  4. 7 kHz
A

6 kHz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Which of the following are used to eliminate doppler mirror image artifact?

  1. Decrease doppler gain
  2. Decrease incident angle
  3. Scan more perpendicular
  4. Decrease the sample volume size