Rapid Loss of Chamber Pressure (IL)
A
- Add gas to maintain and regain original depth (ALP-V-5)
- Shut IL slow and fast exhaust (slow EXH-V-5, fast EXH-V-6)
- Shut IL exhaust hull stop valve (EXH-V-7)
- Shut IL pressure relief hull stop valve (ATM-V-9)
- Divers check disconnect and listen for BIBS leaking
- Shut IL BIBS exhaust hull stop valve (EXH-V-2)
- Check depth (if depth gauges are leaking shut the depth gauge stop valves)
- Check door gaskets and pipe systems
- Check shut med lock pressurization hull stop
2
Q
Rapid loss of Chamber Pressure (OL)
A
- Use ALP-V-13 to maintain depth
- Check shut fast exhaust (EXH-V-8)
- Shut OL exhaust hull stop valve (EXH-V-9)
- Shut OL BIBS exhaust hull stop valuable (EXH-V-4)
- Check shut OL door equalization stop valve (EQ-V-4)
- Instruct tender to check shut OL door equalization stop valve (EQ-V-5)
- Divers disconnect OL BIBS and listen for leaking
3
Q
Increase in Chamber Pressure (IL)
A
- Check shut/shut pressurization Control Valve (ALP-V-5)
- Open IL fast exhaust as necessary to maintain depth (EXH-V-5)
- Divers disconnect BIBS and listen for leaking
- Shut BIBS O2 hull stop valve (OLP-V-6)
- Shut BIBS air hull stop valve (ALP-V-11)
- Shut mix gas stop valve (MLP-V-4)
4
Q
Increase in Chamber Pressure (OL)
A
- Check shut/shut pressurization Control Valve (ALP-V-13)
- Diver disconnect BIBS and check for leaking
- Shut BIBS air hull stop valve (ALP-V-15)
- Shut BIBS O2 hull stop valve (OLP-V-8)
5
Q
Contaminated Atmosphere (IL)
A
- Don BIBS ** tender dons mask first
- Shut primary air shut off valve (ALP-V-3)
- Open secondary air shut off valve (ALP-V-9)
- Preform a hurricane vent as determined by the dive supervisor
6
Q
Fire in Chamber (IL)
A
- Shut BIBS O2 hull stop valve (OLP-V-6)
- Shut BIBS mix gas hull stop valve (MLP-V-4)
- Divers connect BIBS to air QD fitting and don BIBS
- Turn off ECS and CO2 Scrubber
- Divers extinguish the fire
- Stand by for hurricane vent as directed by the dive supervisor
7
Q
Fire in Chamber (OL)
A
- Shut BIBS O2 hull stop valve (OLP-V-8)
- Divers connect BIBS to air QD fitting and don BIBS
- Divers extinguish the fire
- Stand by for hurricane vent as directed by the dive supervisor
8
Q
Contaminated Atmosphere (OL)
A
- Chamber Occupants go on air BIBS
2. Standby for hurricane vent as directed by dive supervisor
9
Q
Loss of Oxygen
A
- Diver removes BIBS mask from patient
- IT checks for proper QD connection
- Check O2 supply low pressure gauge for the proper range of 75-100 psi (OLP-G-1)
- Shut, in use, BIBS O2 supply valve (OLP-V-2 or OLP-V-4)
- Open, standby, BIBS O2 supply valve (OLP-V-2, OLP-V-4)
- Check O2 supply low pressure gauge for proper range of 75-100 psi (OLP-G-1)
- Resume O2 treatment as directed by diving supervisor
- Rack determine the cause of O2 loss
10
Q
Loss of Primary Air
A
- Shut primary air shut off valve (ALP-V-3)
- Open secondary air shut off valve (ALP-V-9)
- Maintain depth with primary pressurization valve (ALP-V-5)
- Rack man investigate the cause
11
Q
Loss of Primary Power
A
- Verify UPS has taken over
2. Secure CO2 scrubber and ECS in the chamber - as needed
12
Q
Loss of Mixed-Gas
A
- IT removes BIBS mask from patient
- IT checks BIBS QD for proper connection
- Check Mixed- Gas Pressure gauge for proper range of 75-100 psi (MLP-G-1)
- Close Mixed-Gas supply valve (MLP-V-2)
- Rack Man identified cause of failure