Sentimag Flashcards

1
Q

Yellow counts mean…

A

A positive signal in which the Sentimag unit and probe will provide audible feedback

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What factors can cause positive signals?

A

Magseed Other ferrous metals i.e. instruments Electrocautery Environmental factors

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Red counts represent…..

A

A negative signal in which the sentimag unit and prove will not provide audible feedback

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What can cause a negative signal?

A

Body (water and tissue) Nonferrous metals Electrocautery Distance past the seed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

When to balance?

A

Option 1- Balance the probe each time Option 2- Knowing when to balance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How long does it take the Sentimag unit to warm up?

A

20 minutes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What are the three different balancing techniques?

A
  1. Air 2.Skin 3. Palpate
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

How long does it take to balance?

A

3 seconds

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Probe check requires the sensitivity to be on what?

A

Set on 3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

During the probe check the Sentimag unit should provide a negative reading at what range?

A

270-330(+/-10%)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Which scanning method is used to determine the location of Magseed?

A

Joystick

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

The Sentimag probe should not be placed within _____mm of any part of an operating pacemaker.

A

15mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

True or False: The Sentimag probe is not suitable for autoclaving or disinfection using formaldehyde.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the sensing zone for the Sentimag probe?

A

3 cm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

True or False: The physician has the ability to use their current instrumentation during a procedure.

A

True but the instruments must be removed from the cavity during scanning with the Sentimag probe.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

When bracketing what is the minimum distance between each seed?

A

2 cm

17
Q

What is the size of the Magseed?

A

Smaller than a grain of rice 1x5 mm

18
Q

What is the length of the deployment device?

A

7 cm 12 cm 20 cm

19
Q

How large of an MRI artifact does the Magseed leave?

A

Around 4 cm.

20
Q

Currently how long can the Magseed be implanted for?

A

30 days.

21
Q

Can I place a Magseed in lymph nodes?

A

Not according to the IFU.

22
Q

What is the applicator gauge size for Magseed?

A

18 gauge.

23
Q

What material is the Magseed composed of?

A

Surgical Stainless steel with less than 1% nickel Bonewax is used as a teminal plug in the applicator (the same as RSL).

24
Q

What minimizes migration of Magseed?

A

The woven design minimizes movement by promoting tissue growth.

25
Q

What imaging modalities is Magseed visible under?

A

Ultrasound, mammography, tomo, and Stereo.

26
Q

True or False: The Magseed signal is weakened by tissue type and degrades over time?

A

False.

27
Q

Can the Magseed be used in conjunction with a wire?

A

Yes as long as the wire is titanium.