Sentimag Flashcards
Yellow counts mean…
A positive signal in which the Sentimag unit and probe will provide audible feedback
What factors can cause positive signals?
Magseed Other ferrous metals i.e. instruments Electrocautery Environmental factors
Red counts represent…..
A negative signal in which the sentimag unit and prove will not provide audible feedback
What can cause a negative signal?
Body (water and tissue) Nonferrous metals Electrocautery Distance past the seed
When to balance?
Option 1- Balance the probe each time Option 2- Knowing when to balance
How long does it take the Sentimag unit to warm up?
20 minutes
What are the three different balancing techniques?
- Air 2.Skin 3. Palpate
How long does it take to balance?
3 seconds
Probe check requires the sensitivity to be on what?
Set on 3
During the probe check the Sentimag unit should provide a negative reading at what range?
270-330(+/-10%)
Which scanning method is used to determine the location of Magseed?
Joystick
The Sentimag probe should not be placed within _____mm of any part of an operating pacemaker.
15mm
True or False: The Sentimag probe is not suitable for autoclaving or disinfection using formaldehyde.
True
What is the sensing zone for the Sentimag probe?
3 cm
True or False: The physician has the ability to use their current instrumentation during a procedure.
True but the instruments must be removed from the cavity during scanning with the Sentimag probe.