SEE EQUIPMENTS/TECH : Delivery and Airway Equipment Flashcards
What is the volume of the gases that cannot be accurately estimated from the cylinder pressure gauge?
Nitrous
With nitrous the pressure will remain constant until
constant until the LIQUID portion of the gas is depleted.
If you’re doing a case and you suspect a pipeline cross over ,2 corrective actions that must be undertaken are?
Open the backup oxygen cylinder
Disconnect the pipeline gas
When you suspect a pipeline crossover, you disconnect the pipeline gas from the wall why?
Because the anesthesia machine will continue to draw from the higher pressure system even if the lower pressure O2 cylinder is open
If the flow of O2 is 2L/min , and the cylinder’s gauge pressure is 500 psi, how long will it last?
87 minutes
Molecular weight of N2O is
44g/mol , one mole at STP is 22.4 L
Names 3 components of the low pressure system
Common gas outlet
Flow meters
Vaporizer
Flush valve is part of which system?
Intermediate pressure system
Hanger yoke and cylinder pressure gauge is part of what system?
High pressure system
What is the purpose of the Diameter index safety system (DISS)
Prevents incorrect hose attachment.
The DISS Is required by the
American Society for TESTING and MATERIALS for an anesthesia workstation
What is the purpose of the pressure regulator on the gas machine?
Reduces cylinder pressure to a lower pressure before it enters the flow valve.
The pressure regulator decrease the variable cylinder pressure to
45-47 psig, to prevent dangerous levels to enter the flow valve.
What device reduces the pressure of gas from an E-cylinder to a usable pressure?
Regulator
A dula hanger yoke system hold 2 oxygen e-cylinder. If both cylinders are open , where will the gauge indicate the pressure?
The cylinder that has the highest pressure.
The HANGER Yoke functions?
Orients cylinders
provides a gas tight seal
Ensures unidirectional low.
The hanger yoke has a _____system
PIN index systme, in order to prevent the connection of a cylinder of one gas to a yoke intended for another.
What type of valve is an oxygen regulator?
DIAPHRAGM
Describe the design of the THORPE TUBE.
Gas specific
Tapered
Narrowest part of the bottom
Largest part on top
What is the minimum oxygen flow rate through the oxygen flow valve when the anesthesia machine is turned on?
150 ml/min
Oxygen flow valves are designed to deliver a minimum flow of
150ml/min when the anesthesia machine is on
Flow meter knobs are ____coded and _____coded
Color and touch
Ball types floats are read in the
Center of the float
Plumb types floats are read
At the top of the float