Safety Flashcards

1
Q

Who are the main regulatory bodies in MRI safety? (4 marks)

A

HPA

ICNIRP

MHRA

IEC

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What are the dangers from MRI and why? (15 marks)

A
  • Main B0 field = projectile risk and implanted devices such as pacemakers
  • B1 field = RF heating of implanted devices and patients with thermo-reg issues, and if patients form loops with their arms and legs
  • Acoustic noise from switching the field gradients
  • Peripheral nerve stimulation can cause pain due to rapid gradient switching.
  • Asphixation and cold burns from a quench,
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Who publishes guidance on local rules and safety? (1 mark)

A

MHRA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly