Routine Radiography QC Flashcards
Level A Tube and generator QA
1-2 monthly X ray light field alignment Light field Bucky centering Light beam diaphragm calibration Distance / scaling checks Radiation output: - repeatability (consistency) - reproducibility (variation with kV/mA/t, focal spot)
Level B X-ray tube + Generator QA
Radiation output - repeatability and reproducibility Exposure time Tube potential (kV) Cut height Angle of swing
Level A CR QA
1-2 monthly Detector dose indication monitoring Image uniformity Review condition of the plates Low contrast sensitivity (TOR/CDRAD) Limiting spatial resolution check (bar pattern)
Level B CR QA
Annual DDI repeatability Measured uniformity Threshold contrast Erasure cycle efficiency Scaling error - lead ruler check Dark noise (mean pixel value in ROI)
Level A QC DR Digital
1-3 monthly DDI monitoring Image uniformity 4-6 monthly Low contrast sensitivity Limiting spatial resolution
AEC
Sensitivity
Operation of the guard timer
Level B QA DR digital
Annual DDI repeatability and reproducibility Measured uniformity Threshold contrast details Limiting spatial resolution Uniformity of response Scaling error Dark noise
AEC
Repeatability
Reproducibility
Image receptor dose
Consistency between chambers
Level A fluoroscopy QA
1-3 monthly
Dose rate under AEC
Display monitor setup - greyscale stepwedge check
Limiting spatial resolution - huttner / TOR18FG
Threshold contrast TOR 18FG
Radiation and image field size comparison / virtual collimation - alignment of x-ray field with visible field on the monitor
Level B fluoro
Annual
ESD dose rate AEC phantom (20cm water equivalent)
Dose rate at input face of detector (comp against MHRA report)
video voltage
Limiting spatial resolution
Threshold contrast GS2 20cm water phantom
Image res uniformity - visual - mesh object
Level A mammography QA
AEC consistency - post exposure mAs Accuracy of stereotactic Monthly AEC check with 7cm PMMA High contrast resolution / threshold contrast - TORMAM with 4cm PMMA phantom
Level B mammography QA
Mean glandular dose check for the standard breast (4.5cm PMMA)
X-ray / light alignment
X-ray film/detector alignment
Compression check - Newton meter / scales
Tube potential - KV meter
Radiation output reproducibility
X-ray field non uniformity - flat field assessment
AEC compensation - 7 cm PMMA
High contrast resolution / threshold contrast - TORMAM
CT Level A QA
Daily
Warm up check
Im noise - std central 40% ROI
CT number check - for water (0) and high contrast material
Scan plane localisation - check alignment of lights with markers
Scan plane localisation from SPR - compare internal and external lights
Table top travel - move table a known distance and measure travel - compare against the display
CT Level B QA
Annual
IM noise - head and body phantom
CT Number values - catphan
CT number uniformity
High contrast spatial resolution - MTF bar phantom
CTDI - CTDI100 in air on-axis head and body
CTDIvol single slice/rotation - 16cm head 32cm body phantom
Irradiated slice thickness - gafchromic film
Im slice thickness z sens profile
Tube to detector alignment - cor
Documentation for DR QA
IPEM 91
IPEM 32 series.
IPEM 81 - critical exam guidelines