REVOLVE ST TROUBLESHOOTING SECTION
Troubleshooting
Product Features
Marking after heavy bleeding Revolve
ST
Product Features
ST
Product Features
Transport to pathology revolve
ST
Product Features
ST
Product Features
Is there a vacuum error when canister is cracked?
ST
Product Features
Not normally, fluid swirls like a tornado or gets really foamy.
ST
Product Features
No Tissue after biopsy, meds were placed through the stopcock revolve
ST
Product Features
ST
Product Features
Probe attached incorrectly revolve
ST
Product Features
ST
Product Features
Vacuum error during procedure revolve
Product Features
Product Features
Heavy Bleeding with revolve
ST
Product Features
ST
Product Features
Troubleshoot: no saline on initialization.
Product Features
- Check tubing connection (all lines)
2 Ensure saline spike it inserted correctly
- Ensure lines are not crimped or kinked
Product Features
No tissue after biopsy no meds through stopcock
Product Features
- Is collection chamber aligned to the 13th chamber?
- Is the canister cracked?
- Is the white vacuum cartridge inserted correctly?
- Is saline flowing?
- Is the aperture completely covered by breast tissue?
Product Features
Marker difficult to Deploy with revolve
ST/US
Product Features
- Is the aperture open
- Is variable aperture setting at full
- Is alignment correct for the probe and marker applicator
ST/US
Product Features
Why won’t my demo probe load?
ST/US
Product Features
Cutter must be open a slight amount at distal end. This allows for forward motion while loading the probe. If not open use gear on bottom of probe to open a few mm.
ST/US