Record Keeping Flashcards
what information is recorded when using a WD
date
operator
cycle start time
cycle number
wash temperature
disinfection temperature
cycle duration
wash detergent added
did machine indicate cycle complete
what would a fault register include
did any faults occur
if yes describe faults
corrective action
did cycle complete successfully
what is parametric release of loads
release of a batch of sterilised items based on data from sterilisation process
what information needs to be recorded for the steriliser
date
time started
operator
cycle number
sterilisation temperature
daily tests performed
bowie and dick
sterilisation successful
machine identified cycle completed successfully
what is the ACT
automatic control test done weekly
what information is included in the WD ACT
temperature for each stage
time each stage active for
disinfection start and end with stage duration
volume of chemical dosed
cycle time overall
what information is included in the sterilisation ACT
pressure
temperature
result of steam penetration test
air leakage test
air detector function test
where can information on daily and weekly testing be found
SHTMs
where is information collected
recorded in a dedicated logbook for each piece of equipment
what is the value of accurate records
will be audited
diagnostic tool - identify issues
traceability of instruments if issues occur