Recent advances in pediatric CPB / Test 3/3 Flashcards
The primary goal of improving CPB has been ?
Smaller, Simpler, Safer, and more biocompatible.
What is the name of the 1st oxygenator that specifically designed for the neonatal patient and became available in the US in 2007 ?
Sorin D100
Allowed very safe minimum operating levels of
as low as 10mL.
What did the study of the D100 show ?
Required 37% less prime volume Resulted in: Less blood transfusions. Less ventilation time. Less post operative bleeding
An impressive feature of the new oxygenator was that it could adequately oxygenate blood at high flow rates of up to ?
This translates on being able to use these oxygenators in patients as large as ?
2 L/m
15 k/g
What is the name of the first oxygenator by TERUMO that has an integrated ALF ?
Capiox FX Oxygenator
When using oxygenators with integrated ALF, what can we expect?
Lower inotrope score
Decreased ventilation time
Decreased hospital stay
Mast mounted pumps can lead to significant reductions in what ?
They also have an improved ability to safely perform CPB on patients as small as ?
↑Hct.
↓ Prime Volume
↓ Blood Use
3.2 kg
Tubing can account for approximately ___ of the prime volume of neonatal CPB circuits?
3/4
What are the primary uses of a hemoconcentrator (UF) device ?
removal of extravascular water
removal of LMW solutes
↑Hct
Concentrate platelets & plasma Using low prime volume devices translates into what ?
more blood in the venous reservoir
Current trand is to shift from high K+ depolorazing based CLP to ?
Hyperpolarizing or modified polarizing CLP that offers:
Similar myocardial protection
Longer dosing intervals
Reduction in circuitry complexity
Custodial-Histadine-Tryptophan-Ketoglutorate =
Transplant preservation solution
Del Nido Formula ?
1:4 hyperpolarizing blood based solution
Administered at 4 - 6 degree celcius.
Patients 40 kg = 1 L of CLP is MAX dose
Hemochron Jr stats ?
Membrane surface area :
Prime Volume:
Max Ultra Filtration rate:
Membrane surface area : 0.09 M2
Prime Volume: 8 ml
Max Ultra Filtration rate: 30 ml/min
Hemochron Jr mini ?
Membrane surface area :
Prime Volume:
Max Ultra Filtration rate:
Membrane surface area : 0.07 M2
Prime Volume: 14 ml
Max Ultra Filtration rate: 12 ml/min