Real Ear Coupler Difference (RECD) Flashcards
What is RECD?
A sound generating transducer produces a signal in the ear canal and in a 2cc coupler to measure the resonance of each.
RECD is the difference in decibels across frequencies, between an ear canal resonance and the resonance of the 2cc coupler
Why is RECD done?
Conversion from HL to SPL
Difference becomes conversion factor
Used to make coupler simulate the real ear
Can do the fitting and the measurements in the test box without the patient there
Mainly done for pediatrics
Is the RECD different patient to patient?
Yes
What are the two reasons for RECD measurement?
Accurately converts an individual’s HL audiometric thresholds, measured using inserts, to dB SPL values (creates personalized conversion values)
The prediction of real ear output when hearing aid measurements are made in the test box (used to preprogram device)
Why do you want to create personalized conversion values?
Allows you to create accurate fitting targets
Supplies precise adjustments for differences in canal volume and impedance variations
What is the ANSI standard coupler for RECD?
HA-1
What is the ANSI standard for RECD?
RECD measurement can be made with a custom earmold or EAR insert
The same coupler for BOTH measurements (either a foam tip and the HA-1 coupler or a custom mold and an HA-1 coupler)
Why is RECD gross?
Need to use funtak on every device (cannot be standardized)
What was done to make RECD less gross?
ANSI established standard conversion factors between HA-1 to HA-2 coupler to support use of the HA-2 coupler (don’t need funtak)
How is RECD measured?
Signal: Pink noise
The same acoustic signal must be used for BOTH measurements
The difference between these 2 resonances becomes a conversion factor
The conversion factor is added to 2cc coupler resonance to simulate the patient’s ear canal resonance in the 2cc coupler
What is the first step of RECD?
Measure the coupler response using RECD transducer
Navigate to “ON EAR” measures to RECD to “coupler calibration”
What is the second step of the RECD?
Measure the REUR first and then measure the canal resonance using a foam insert and the insert transducer
Use Aquaphor on the insert or earmold to reduce the slit leak
What does a negative LF value suggest in the RECD measurement (not the difference)?
Slit leak present
Negative LF results are expected when perfs or PE tubes are present (volume is larger)
How do you correct a slit leak?
Make sure you allowed the insert to fully expand
Increase the tip size
Apply Aquaphor to the foam tip to improve the seal
What does a negative RECD (>10 dB) in the 4-6 kHz region suggest?
Suggests the probe tube is blocked
What are the verifit 2 protocol differences for RECD?
Called WRECD (0.4cc coupler used to mimic the size of the ear canal)
What should you expect on the verifit 2 (WRECD)?
A negative LF responses on adult ears - a very large negative slope in the LF may still indicate the presence of a slit leak
Ear canal resonance above 1k Hz is equal to but not below the 0.4 cc coupler response
If it falls below the WRECD above 1 kHz it still indicates a blocked or pinched probe tube
When the RECD is above the dotted line, what does that mean?
The ear canal is smaller than normal
When the RECD is below the dotted line, what does that mean?
The ear canal is larger than normal