Random Questions Flashcards

1
Q

Adequate number of electrodes are essential to ensure?

A

To ensure the eeg activity having a small area of representation is recorded & To analyze accurately the distribution of more diffuse activity.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

A small number of electrodes (<16) is appropriate for normal EEG recordings. True or False

A

False

Why? 21 electrodes covers the surface area, 16 channels.

(19 plus ground and reference totals 21 electrodes).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Ground electrodes in EEG in digital EEG devices serve as a safety hazard just as were earth grounds in early analog EEG devices?

A

False : Ground electrode is earth ground is built in amplifiers works different and not needed.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

No electrode should ever be connected to the _________ of the equipment or the ____________.

A

Chassie , earth ground

Electric hazards

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

__________ should be checked as a routine pre-recording procedure during EEG recording.

A

Inter-electrode impedances under 5 ohms

Do not over prep to cause skin breakdown or break skin.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Impedances that are balanced and up-to 10kOkms are acceptable for EEG recordings.

A

True

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Impedances are important for common mode rejection.

A

True

Common mode rejection: is the amplifier the takes in the same noise, filtered same noise and then the amplifier amplifies the difference of the noise.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Short circuit and shunt is indicated on the EEG by ________ and possibly related to ______ in the scalp.

A

Impedances below 100 Ohms & Salt bridge

Related to high and unbalanced Ohm -flatline - re-prep location.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Recording electrodes should be free of ______ & _______ and should not significantly attenuate signals between _____ Hz ____ Hz.

A

Inherent noise, Drift
0.5Hz, 70Hz

Channels are high quality and calibrate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Meaning of IFCN is

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Meaning of IFCN is

A

International Federation of Clinical Neurophysiology.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Meaning ACNS

A

American Clinical Neurophysiological Society.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Beyond consideration of patient discomfort and risk of injuries to personnel, needle electrodes are disadvantageous because?

A

They have higher impedances which results in potentially higher levels of noise.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Reference electrodes is one of the 10/10 or 10/20 system electrodes? True or False

A

False

It is a required electrode, they are not in the placement no specific location on the head.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

The use of link ear as a digital recording reference is specifically encouraged.
True or False

A

False

linked ear montage can not produce accurate reading.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Contra -opposite side
Epsi- same side

17
Q

_________ calibration is not necessary for digital system but may occasionally help in detecting errors in the montage selection process for analog recordings.

A

Biologic

System reference replaced the Bical eeg.

18
Q

_________ is defined as the ratio of input voltage to trace deflection.

A

Sensitivity

19
Q

How many minimum channels required for recording the EEG to decrease error as well as for transient activities?

A

16 Channels additional channels are often needed for monitoring other physiological activities and for highlighting activity recording in specific regions

20
Q

______ wiring should meet the underwriters Laboratories standards required for hospital ser-vice, as per state and local electrical code.

A

AC current

21
Q

According to ACNS guidelines what is the settings preferred for recording EEG (HFF, LFF, Notch Filter Sensitivity).

A

LFF 1Hz, HFF 70Hz, Notch 60/50 Hz. Turned off always, sensitivity 5-10uV/mm start at 7uV/mm

22
Q

Is the time constant the same as the Low frequency filter.

A

They are not. They are very different.

Time constant is a 0.6 or 0.3. Very low

23
Q

What is Neuron?

A

Building blocks of nervous system

24
Q

Human brain has how many neurons?

A

60 billion neuron

25
Q

Presynaptic neuron sends signals to the postsynaptic to become more active

26
Q

Presynaptic sends negative to post synaptic signals

27
Q

Cerebral Hemisphere

A

83% of total brain mass, covers the diem cephalon and the upper brain stem
Bumps (guru) and grooves (Sulci)
Characteristic patterns of sulci fissures over surface.

28
Q

Korbinian Brodmann 1868-1918
German Neurologist
Died 50 yrs

A

52 cortical areas the differs histologically