RADIOGRAPHIC ANF FLUOROSCOPIC FEATURES Flashcards
RADIOGRAPHIC PROTECTION FEATURES
Protective Xray Tube Housing
Control Panel
Source-to-image-receptor distance indicator
Collimation
Positive-Beam Limitation
Beam Alignment
Filtration
Reproducibility
Linearity
Operator Shield
Mobile xray imaging
Purpose of Protective Housing
Reduces leakage (scatter) radiation
Leakage radiation must be less than __ at a distance of __ from the protective housing.
100 mR/hr (1 mGya/hr); 1 m
Must indicate the conditions of exposure and must positively indicate when the x-ray tube is energized.
Control Panel
What signals indicate when the x-ray beam is energized?
visible or audible signals
X-ray beam on must be __ and __ indicated to the radiologic technologist.
positively; clearly
can be as simple as a tape measure attached to the tube housing, or as advanced as lasers.
Source-to-image-receptor distance indicator
SID indicator can be as simple as a __ attached to the tube housing, or as advanced as __.
tape measure; lasers
SID indicator must be accurate to within __ of the indicated SID.
2%
Light-localized, variable-aperture rectangular
Collimation
may replace the collimator for special examinations.
Cones and diaphragms
Attenuation of the __ by collimator shutters must be equivalent to attenuation by the __
useful beam; protective housing
The x-ray beam and the light beam must coincide to within __ of the SID.
2%
Automatic, light-localized, variable-aperture collimators were required on all
Positive Beam Limitation
Special x-ray imaging systems manufactured in the United States between __
1974 and 1994.
devices are no longer required but continue to be a part of most new radiographic imaging systems.
positive-beam–limiting (PBL)
The PBL must be accurate to within of the SID
2%
mechanism to ensure proper alignment of the x-ray beam and the image receptor.
Beam Alignment
total filtration (inherent plus added) of at least __ when operated above __ kVp.
2.5 mm Al; 70
Radiographic tubes operated between __ must have at least __.
50 and 70 kVp; 1.5 mm Al
Below __ kVp, a minimum of __ total filtration is required.
50; 0.5 mm Al
X-ray tubes designed for mammography usually have __
30 µm Mo or 60 µm Rh filtration.
should be constant from one exposure to another
output radiation intensity
Reproducibility is checked by
making repeated radiation exposures through the same technique and observing the average variation in radiation intensity.
variation in x-ray intensity should not exceed __.
5%