Rad physics, equipment, apps. Flashcards
What is a grid? (made of, purpose, when do you use it)
A Board made of lead strips with radiolucent material separating the strips. Purpose is to reduce scatter from reaching the IR. Increase contrast & image detail and reduce image degradation.
Use when using more than 70 KVP, and when the patient thickness is more than 10cms.
What is grid ratio?
height of the strips divided by the width of the interspace material.
What is grid frequency?
Number of lead strips per centimetre.
What is grid factor?
What we X the mAs by so that quantity of XR beam is enough to penetrate the grid, IR & pt.
What is scatter?
Radiation that is formed by the interaction of the primary beam with atoms in the object being imaged. Influenced by pt thickness, kVp, collimation.
What type of grids do we have?
What grid ratios do we have.
Bucky= focussed grid. Have a specific SID, centered to grid, allows for more transmitted photons to reach the IR. They are oscillating, helps blur out grid lines. Solanoid gives the grid a tug, release and is linked to exposure button.
110cm Grid (109-140cm). Grid ratio is 12 : 1 x 5.
140cm Grid (109-210). Grid ratio is 8 : 1 x 4. (cleans up 85% of scatter).
What Is grid cut off?
When radiation is absorbed unevenly across film, misuse of a grid.
Peripheral grid cut off
Absorption of the primary beam due to the divergence of the beam at the periphery of a parallel grid
Off focus grid cut off
Peripheral grid cut off that occurs because the SID is not within the reccommend focal range. Higher grid ratios have narrower focal ranges.
Off level grid cut off
decrease in density across the film that occurs when the grid or tube is angled.
Off centre grid cut off
Decreased density across the film caused by incorrect centering. The center of the grid must be positioned directly under the x-ray tube target. Correct centering is more important with higher grid ratios.
Upside down grid cut off
Severe cut off on either side of the central ray and increased density in the middle of the film caused by placing a focused grid in upside down.
As grid ratio ___, radiographic density ___.
As grid ration increases, radiographic density decreases.
Explain the air gap technique.
Used instead of a grid as the scatter radiation is attenuated in the air between the patient and the IR before it reaches the IR. Hence, the air is acting as ‘a grid’.
Because of the large OID, detail of the image is lost, therefore SID must be increased from 110cm to 180cm.
Explain KVP
Quality of the beam penetration, higher the penetration, higher the KVP. Amount of energy of the xray is determined by the voltage applied from cathode to anode.
mAs
Quantity. The amount of xrays reaching the IR (produced), directly related to patient dose. Energy is not changed, just the number. Amount of xrays is determined by the current.
Explain the MAS button…
Goes up by 25% of original mAs. Must go up 25% to see a noticeable difference.
Explain the density step button…
Goes up in 1/2 steps (12.5%), this gives you 12.5% more time before the AEC cuts out. Must go up two 1/2 steps to increase time to 25%. 25% change- noticeable difference.
Explain the body habitus button…
Preprogramed= depends on body type/ pre-set. Goes up in KVP, density step stays the same, but mAs goes up according to the pre set, more dose to patient.