rad phys questions Flashcards
to double the density of a film by using mAs what would you do?
increase the mAs by 100%
of the choices below which is true about a reciprocating grid?
a. requires a 10:1 ratio
b. it requires 103 lines per inch
c. it is thicker than a stationary grid
d. it requires a faster exposure time
e. it is less expensive than a stationary grid
c. it is thicker than a stationary grid
when converting from a grid with low ratio to one with high ratio which of the following would be the best to do to maintain a good film quality?
a. increase kV
b. increase mAs
c. decrease kV
d. decrease mAs
b. increase mAs
which of the following is recommended for spinal radiography in the usual chiro office?
a focused grid
what effect does using a high kV have?
produces more scatter
more Compton compared to photoelectric effect
more penetration= less absorption
more graying from film fog
more x-rays pass through without interaction
what is the preferred grid focus range for a spinal radiography?
40” to 72”
how do you control scatter from a high kV?
use the optimal kV technique
what is the relationship of beam size and scatter?
the larger the beam the more scatter
what do you do to control the scatter from a large size beam?
collimate to film or part size
describe optimal kV
the lowest possible mAs with acceptable contrast on the film
how many lines per inch are recommended for a stationary grid?
100-103
diminished penetration at the periphery of a grid is called?
grid cut off
if a film is said to be burned out, what would it look like?
too dark
which choice below is the main reason why a “fixed kV technique is preferred?
a. contrast is consistent
b. less stress on the tube
c. patient exposure is reduced
d. faster exposure time can be used
a. contrast is consistent
scattered or primary rays that leave the patient and reach the grid is called?
remnant radiation
what are some common grid ratios and there max kV?
8: 1- 90 kV
10: 1- 100 kV
12: 1- 110 kV
16: 1
which grid ratios are recommended for spinal use?
10:1 or 12:1
grid frequency is the # of lead strips per in also called?
lines per inch LPl
what are the 3 types of stationary grids?
crossed- not practical for spinal
parallel
focused
with parallel strips there is an attenuation of the periphery of the beam due to an angle of diverging rays, what does this create?
uneven exposure with sides being lighter
which type of stationary grid has graduated angles to match the angle of divergence?
focused- the periphery of the beam is not attenuated
what is grid radius?
the angle of the strips that match the divergence extending from the grid to the focal spot
which type of grid will produce the least amount of grid lines?
moving grid because moving the grid during exposure can avoid the lines all together
when are grid lines more noticeable?
with lower frequencies and parallel grids
when are grid least noticeable?
with a 10:1 focused grid and 103 lines
what is another name for a moving grid?
reciprocating grid or Potter Bucky diaphragm
exceeding the optimal kV while still maintaining good density will yield what type of film?
long scale contrast
which of the following would increase patient exposure?
a. using high kV
b. using 10:1 stationary grid
c. using high speed screen
d. using air gap technique
b. using 10:1 stationary screen
about how much increase in mAs is required when going from a non grid technique to using a technique with a 12:1 grid?
12 times
what are some negatives to using a moving grid?
more expensive
can fail
thicker so increase OFD
can cause motion of film leading to blurring
can be too slow for fast exposures causing grid lines