Quiz3 Wed Flashcards
Spin top test (half wave)
•time x 60= # of dots
Spin top test
- disk with hole exposure is made while top is spinning
* used to test timer
Spin top test (full wave)
Time x 120= # of dots
Spin top (three phase)
- can’t be used
* must use synchronized timer
Wire mesh tool
•used to check for screen film contact
Does the temperature of the developer effect density?
- yes
- to hot= too dense “overexposed”
- to cold= less dense “underexposed”
Processing artifact which looks like lightning or trees?
•static- static discharge from low humidity
Field size increases as
•SID increases
If distance is halved then mAs is
•1/4 the original
Geometric unsharpness is due to
- focal spot
- SID
- OID
Subject unsharpness is due to
- patient thickness
- tissue density
- object shape
Xray & light field congruence
- must be within 2% of SID
* test tool or coin is used if off adjust mirror
If SID is outside the grid radius what is the result
•grid cutoff
Reproducibility
•Output radiation intensity should be constant from
One exposure to the next
•should not vary more than 5%
Linearity
- same mAs settings with different mA & time should be same output
- should not exceed 10%