Questions Flashcards
high bone marrow examinations
- barium enema
- upper GI series
- abdominal angiography
biological effects of ionizing radiation that are manifested in offspring
genetic effects
cell line used as an indicator of somatic effects of radiation
bone marrow
photocathode gives off what in proportion to the intensity of fluorescent light?
electrons
in an x-ray system with ABC and an x-ray tube BELOW the table, moving the image intensifier AWAY from the patient will do what to the patient’s radiation exposure?
increase
cell type most sensitive to radiation
lymphocytes
strong correlation between the mean radiation dose received by bone marrow and what?
leukemia
what is half-value layer (HVL)?
thickness of absorbing material necessary to reduce x-ray intensity to 1/2 its original value
collision between an x-ray photon and INNER orbital electron, knocking the electron out of its orbit
photoelectric effect
what organ is NOT used for somatic dose measurements
thyroid
what are required to be posted in x-ray areas?
- form RH-2364 “Notice to Employees”
- copy of operating procedures
tighter collimation to reduce the FOV will cause image brightness to
reduce
minimum lead equivalent thickness for protective aprons
0.5 mm
probability of an effect occurring
stochastic effect
severity of an effect that has occurred
nonstochastic effect, aka, deterministic effect
in digital imaging, adjusting the window level changes
contrast
photoconductive target function
allows an electron charge to be generated
can operator exposure occur after the dead-man switch is released?
NO
hypothesis states that any radiation dose has some degree of effect
non-threshold
type of sterility resulting from a 50 rad exposure to gonads
temporary
reason filters are used in film badges
measure doses over a wide range of energies
type of image receptor converts x-rays into an electrical charge using an array (matrix) of detector elements
flat panel detector
bucky slot cover lead equivalent thickness
0.25 mm
dose equivalent limit to an embryo/fetus during the ENTIRE pregnancy for an occupational exposure of a pregnant woman
5 mSv
image characteristics greatest at the center of the image intensifier
- resolution
- detail
- brightness
an operator should use a thyroid shield when
performing fluoroscopy and standing close to the patient
type of image receptor uses either direct or indirect methods to convert x-rays into an electrical charge
flat panel detector
conversion factor formula
intensity of output phosphor ÷ mrads/s
term refers to the standardized method of evaluating the brightness of an image intensifier
conversion factor
closed-circuit systems transmit signals through
cables
component of a fluoroscopic system that speeds up electrons
accelerating anode
when the fluoroscopy time limit has been reached, the manual-reset timer must do the following
- produce an audible signal
- interrupt the x-ray beam
minification gain x flux gain =
total brightness gain
occurs when the diameter of the image from the optical system is SMALLER than the shortest dimension of the film
underframing
mild adverse effects from contrast media
- mild nausea/vomiting
- limited urticaria
- limited “itchy” throat
- nasal congestion
- sneezing/runny nose
moderate adverse effects from contrast media
- protracted nausea/vomiting
- diffuse urticaria
- diffuse erythema
- facial edema w/o difficulty breathing
- wheezing w/o hypoxia
severe adverse effects from contrast media
- facial edema w/ difficulty breathing
- diffuse erythema w/ hypotension
- laryngeal edema or wheezing w/ hypoxia
- anaphylactic shock
minimum required thickness of protective curtains
0.25 mm Pb equivalent
when should the State be notified If an occupational worker receives an exposure of 0.02 Sv for 3 consecutive calendar quarters?
w/i 30 days
immediate notification to the Radiologic Health Branch is required if an individual has received
- total effective dose equivalent ≥ 25 rems (0.25 Sv)
- eye dose equivalent ≥ 75 rems (0.75 Sv)
- shallow-dose equivalent to the skin or extremities ≥ 250 rads (2.5 Sv)
24-hour notification to the Radiologic Health Branch is required if an individual has received
- total effective dose equivalent ≥ 5 rems (0.05 Sv)
- eye dose equivalent ≥ 15 rems (0.15 Sv)
- shallow-dose equivalent to the skin or extremities ≥ 50 rads (0.5 Sv)
doubling exposure time does what to the total exposure of the patient
doubles
legal aspects of taking X-rays fall under
- civil law
- regulatory law
- administrative law
how many days after the onset of menses is it safest to perform x-rays?
10-14 days
irradiated animals die from
the same diseases as non-irradiated controls but at a younger age
what are the specified parameters to measure fluoroscopy exposure rates using a phantom?
- maximum 5 rads/min
- 30 cm from the image intensifier
- using a phantom 9 in of water
what must be posted in areas where x-ray machines are being used?
- x-ray supervisor and operator permit
- radiologic technologist certificate
- RHD’s “Notice to Employees” (form RH-2364)
curve used as a basis for regulatory radiation guidelines
curve that intersects the abscissa at the origin
if the x-ray tube is fixed BELOW the table in a system equipped with ABC, what effect will moving the image intensifier AWAY from the patient have?
increase the patient dose
recording system that can reduce patient dose up to 95%
video disc recording
of horizontal lines on a television monitor
525
identify the labeled components
I. = photocathode
II. = input phosphor
what is the process in the image intensifier where electrons are accelerated from the photocathode to the output phosphor?
electron intensification
what is the minification gain in an image intensifier with a 10 in input phosphor and a 1 in output phosphor?
100
point where the dose-effect curve intersects the abscissa
threshold dose
fluoroscopic examinations should
- minimize patient exposure
- minimize operator exposure
- obtain necessary diagnostic information
is a dose to the embryo/fetus considered a genetic dose?
NO
which of the following is directly proportional to the exposure delivered to the patient?
mA
how many rads of acute X-radiation in the diagnostic energy range to the eyes will result in cataracts?
200-300 rads
component that controls the radiation exposure incident at the input phosphor in cine film imaging
aperture
term that refers to the frequency range of a video signal
bandpass
radiation damage exhibited in the offspring of people exposed to radiation
genetic dose
disadvantages of using pocket ionization chambers
- sensitive to mechanical shock
- must be calibrated periodically
- do not provide a permanent record
- information is lost if over-exposed
sends light from the image intensifier to viewing and recording systems
camera control unit
half-value layer (HVL) is a characteristic of an x-rays
quality
thyroid shields are used by operators
when the operator is close to the patient during fluoroscopy
structure does light hit first in the human eye
cornea
factors that lead to radiation exposure to the operator
- radiation leakage from the x-ray housing
- scatter off the patient
- scatter off the collimator
- scatter off the table top
maximum monthly radiation dose to an embryo/fetus
0.5 mSv
mutations from radiation exposure tend to be
deleterious
material that absorbs less penetrating x-rays from the useful beam
filter
minimum total filtration required in fluoroscopic equipment at normal operating voltage
2.5 mm Al equivalent
approximate dose to a fetus from an upper-GI examination of a pregnant patient with a total skin exposure of 5 Roentgens
100 mrads
is NOT a known form of radiation-induced cancer
retinoblastoma
predetermined time limit for a cumulative manual-reset timer
5 minutes
personnel monitoring devices that discharge a capacitor by the ionization of air when in the presence of radiation
- pocket chamber
- pocket ionization chamber
cancers that have been seen in humans exposed to high levels of radiation
- leukemias
- bone malignancies
- skin malignancies
what factors will influence the effect of radiation exposure?
- radiation type
- radiation dose
- cell type
formula to calculate lifetime (cumulative) occupational exposure limit
10 mSv x years
time it it takes to change a fluoroscopic system from a vertical to a horizontal imaging position
2 seconds
what happens if the FOV is held constant and the matrix size decreases?
- pixel size increases
- spatial resolution decreases
bone marrow areas typically exposed when wearing a lead apron
- clavicles
- skull
- arms
sperm precursors
spermatogonia
film size in spot filming cameras has what effect on dose to the patient?
larger film size = greater radiation dose
tabletop intensity of the X-ray beam for a system operating at 80 kVp should not exceed how many rads per minute
2.2 per mA
as matrix size increases
- pixel size decreases
- spatial resolution increases
major source of secondary radiation in radiography is from
patient’s body
people that must be monitored regardless of their likely exposure
- people who operate mobile x-ray equipment
- people who enter a high radiation area
reproductive cells that are irradiated may have changes in their
- genes
- chromosomes
visual acuity of rods (scotopic vision) is what compared to cones (photopic vision)?
worse
used to compare the spatial resolution of imaging systems
modulation transfer factor (MTF)
with optical coupling between an image intensifier, and viewing and recording systems, how much light is sent to the recording system (cine camera)?
90%
does NOT affect patient exposure
room lighting
of horizontal scan lines on a television monitor
525
reason film badges are used
- provide a record of exposure
- measure operator exposure
accounts for the majority of the population’s exposure to man-made radiation
medical use
acute dose of radiation that forms cataracts
200 rads
effects from radiation exposure that occur long after exposure are referred to as
- delayed effects
- long-term effects
- late effects
Sievert is a measure of
absorbed dose equivalent
animal experiments have shown that deleterious effects to an embryo can be produced with as little as radiation exposure as
10 rad
resolving power measurement
lp/mm
estimated percentage of the population that are exposed to fluoroscopy yearly
5%
pixels are given values for a certain color or gray level; the number of bits used to encode these pixel values is known as
bit depth
primary purpose of protective drapes
to reduce scatter to the operator
digital fluoroscopy systems are commonly used to record images of
vessels
occurs when the diameter of the image from the optical system is LARGER than the shortest dimension of the film, resulting in part of the image being lost
overframing
total number of cycles per second available for display by the television camera
bandwidth (bandpass)
skin dose is a good indicator of
dose to nearby organs
the following are required to be posted in x-ray areas
- supervisors and operators permits
- technologists fluoroscopy permits
- radiographers certificates
what happens to a patient’s entrance skin exposure (ESE) when the FOV is decreased on the image intensifier?
increases
least radiosensitive cell type
muscle
can be directly linked to long-term radiation exposure
cataracts
primary radiation is considered
useful x-ray beam
output phosphor converts electrons into
light
ratio between the vertical resolution and the number of scan lines
Kell factor
in California, how frequently must tube potential (kVp) and current (mA) must be monitored?
weekly
what is an eye dose equivalent, and what is the annual limit?
- radiation exposure to the lens of the eye
- 0.15 Sv
the output phosphor converts photoelectrons into
light photons
how many mA would be used during spot filming
100 mA
thickness (in lead equivalent) of protective curtains
0.25 mm
grainy appearance in an image caused by statistical fluctuation of absorbed X-ray photons
quantum mottle
has all its signals carried through cables
closed circuit TV system
absorbed dose (rads) can result in spontaneous abortion of a fetus
50 rads
medical use accounts for what percentage of the average population’s exposure from man-made radiation?
90%
when electrons are accelerated from the photocathode to the output phosphor
electron intensification
a higher frame rate sequence leads to
- higher patient dose
- improved motion aspect as seen by the operator
major source of secondary radiation in radiography
patient’s body
what are personnel monitoring devices
- pocket ion chamber
- thermoluminescent dosimeter (TLD)
- film badge
increasing the kilovolt peak (kVp) and decreasing the milliamperage (mA) will have what effect on skin exposure?
decrease
in cinefluorography, term for camera shutter opening at the same frequency as x-ray pulses
synchronization
can humans detect ionizing radiation as it passes through their body?
NO
fluoroscopy technologists can do what?
- report a diagnosis to a patient if ordered by a licentiate of the healing arts
- position patients correctly during fluoroscopy
- check that images are of high enough quality
purpose of an electron gun
to produce a scanning electron beam
factors directly proportional to milliamperage (mA)
- X-ray output
- radiation dose to patient
clinical benefits of PACS
- simultaneous viewing of images
- faster turnaround of diagnosis
- digital images
what is NOT a clinical benefit of PACS?
hard copy backups
maximum kilovoltage peak (kVp) typically used during fluoroscopy
150 kVp
x-ray operator is required to wear personnel monitoring equipment when?
only when working
resolution of modern cesium iodide image intensifier tubes
4 lp/mm