Quality Control Flashcards
1
Q
Dose calibrator geometry
A
Tests for measurement of activity as the volume of a radioactive source changes
Installation and after repair
2
Q
Dose calibrator accuracy
A
To check the ability of the dose calibrator to measure different level gamma energy (100-500 KeV)
Annually
3
Q
Dose calibrator linearity
A
To check ability of the dose calibrator to measure a wide range of activity from mCi to uCi amounts
Quarterly
4
Q
Dose calibrator constancy
A
To check consistency of the source of known activity from day to day
Daily