QC Digital Flashcards

1
Q

Half-value measurement for testing filtration/quality of beam

A

Minimum amount at 70kvp is 2.5 mm of aluminum equivalent

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Light field to radiation field alignment - accuracy of collimator

A

Must be accurate to +- 2% of SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Positive beam limitation (testing collimation)

A

Primary beam should not be allowed to be larger than the IR when activated

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Central ray alignment

A

centering mark should be +- 1% of the light field central ray

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Focal spot size

A

size should be within +- 50% of equipment specifications

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

kVp accuracy

A

should be within +- 3-5 kVp on control panel OR +- 10% of kVp indicated

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Timer accuracy

A

should be within +- 5% for exposure longer than 10 milliseconds
OR +- 20% for exposures less than 10 milliseconds

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

mA linearity

A

should be within +- 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

mA reproducability

A

should be within +- 5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

mA reciprocity

A

should be within +- 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

AEC Ion chamber sensitivity

A

should be w/in +- 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

IR Exposure Variation (density adjustment)

A

vary per unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

AEC response capabilities

A

should be w/in +- 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

backup timer verification

A

backup timer should terminate exposure after prolonged on time (tested by placing lead on the detector and waiting for the backup timer to kick in)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Fluoroscopy exposure rate

A

No more than 100 mgy/min and no more than 10 R/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Automatic Brightness Control

A

Checked to make sure the radiation dose hitting the input phosphor is constant. Use phantoms to measure if dose is adjusted.

17
Q

Plate uniformity

A

Numbers on plate should be similar in value. Less than 20 value difference should be seen amongst the measured areas.

18
Q

Spacial resolution line pairs of digital imaging systems

A

2.5 lp/mm

19
Q

Grayscale standard display function

A

Should not vary more than 10% as recommended by American Association of Physicists in Medicine task group 18

20
Q

Average luminance of primary display monitors

A

300-500 cd/m2

21
Q

How often should aprons be checked?

A

Visually every 6 mths

Checked fluoroscopically annually