TG-66 QA of CT Image: Daily Test
Alignment of gantry lasers with the center of imaging plane +/- 2mm
CT number accuracy for water +/- 5 HU
Image noise
In plane spatial integrity +/- 1mm
TG-66 QA of CT Image: Monthly Test
Orientation of gantry lasers with respect to imaging plane +/- 2mm
Spacing of lateral wall lasers +/- 2mm
Orientation of wall lasers +/- 2mm
Orientation of ceiling laser +/- 2mm
CT number accuracy 4-5 different materials
In plane spatial integrity in x and y direction
Field uniformity most commonly used kVp
TG-66 QA of CT Image: Annual Test
CT number accuracy electron density phantom
field uniformity all kVp settings +/- 5 HU
electron density to CT number conversion
Spatial resolution
Contrast resolution
*CT dose index, patient dose from CT scan +/- 20% manufacturer specifications
Dose from CBCT
1 cGy
AAPM TG142 Imaging QA for the Linac: Daily
Planar kV and MV (EPID) imaging:
AAPM TG142 Imaging QA for the Linac: Monthly
Planar MV imaging (4 cardinal angle coincidence (2mm nonSBRT, 1mm SBRT), scaling, spatial resolution, contrast, uniformity and noise)
Planar kV imaging (coincidence and scaling:
AAPM TG142 Imaging QA for the Linac: Annual
Planar MV imaging full range of travel SDD +/- 5mm; imaging dose
Planar kV imaging beam quality/energy