PRVC and Auto Adjusting Modes Flashcards
What is PRVC on a Servo vent?
PRVC
What is PRVC on a PB vent?
VC+
What is PRVC on an Evita?
CMV or VC with autoflow
What is PRVC on a Hamilton vent?
APV-CMV
What main advantage does PRVC have over pressure control?
The tidal volume can vary greatly in pressure control, but PRVC will target a specific volume without exceeding safe pressures
How does the vent decide what pressure to use in PRVC?
Most vents deliver one VC breath to determine Pplat, then deliver a PC breath using pressures that are approximately the Pplat - PEEP. It then adjusts based on the Vt achieved.
What are the advantages of using PRVC?
- lower ventilating pressures
- breath is more adaptive if patient has drive to breath
- breath is more comfortable
- Vt may vary but MV and average Vt are fairly consistent
- May decrease risk of overdistension
What are some disadvantages of PRVC?
If the patient is changing rapidly the machine may struggle to keep up. If the patient is breathing spontaneously it may decrease driving pressure to the point of exhausting the patient.
What is a PRVC reset?
When there are large changes in Vt the mode will reset and deliver a VC breath to try to measure Pplat again. This often happens in bronchospastic or coughing patients and can make things worse.
What is the PRVC pressure limit?
It is a limit automatically set by the machine below the high pressure alarm that will allow the breath to continue instead of just cycling off
How should we set the pressure alarm in PRVC?
Add the amount of the pressure limit to the current PIP and then set the PIP alarm 5 above that.
What are the pressure limits on the different kinds of vents?
PB: 3 cmH2O
Hamilton: 10 cmH2O
Any other vent: 5 cmH2O
What is the Mandatory Minute Ventilation mode?
As long as the patient’s minute ventilation is higher than what is set, the patient is in PSV. If the patient’s minute ventilation falls below this, they are given control breaths to increase it above the set level.
How do we set the MV in MMV?
Set it lower than the current spontaneous level, but high enough to provide at least adequate ventilation.
Why is MMV not used often?
It’s only available on Evita ventilators, the patient defaults to VC which is not comfortable, the patient is able to maintain adequate MV even when tiring out, and the RR will typically elevate as Vt declines.