Product Training Flashcards

1
Q

Define Seizure Burden

What would 10% seizure burden mean?

A

% of time the patient is experiencing seizure or seizure like activity over the last 5 minutes.

That means that 10% of the time over the last 5 minutes the patient was experiencing seizure or seizure like activity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is our sensitivity for alerting for suspected Status Epilepticus?

A

95% sensitivity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What are the three main objectives for an ICU In-Service

A

Why do you use Ceribell
How does it work
How is it specifically going to work at Hospital X

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

System Setup - How should the patient’s head be supported when setting up the Ceribell EEG?

A

Roll up one large towel or two small towels to support patient’s lower head, to create a space between patient’s head and pillow to prevent pressure from the headband electrodes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

System Setup - What should be done with a patient’s hair when setting up the Ceribell EEG?

A

You should pull long hair behind the patient’s ears and move the hair bangs out of the way

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

System Setup - What is the correct position for the headband? 3 bullets

A

Clasp centered
Right above the eyebrows
Right above the ears

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Can you use the Ceribell headband on patients with head wounds?

A

Yes. You can position the headband to avoid the wound if needed, but you should include a note and notify the physician so that the physician is aware of how the headband was placed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What does it mean when the electrodes on the recorder screen are shown in grey?

A

Cable?****

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the correct way to prepare the electrodes?

A

While applying gentle pressure, rotate the electrodes back and forth for 7 complete rotations, and then use two thumbs to push the plunger

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

I rotated one electrode 7 times and deployed the gel, but the electrode light in the recorder remains red. What should I do?

A

If you have prepared all of the electrodes and the indicator is still red, rotate the electrodes 5 more times until the light turns green. Use a syringe to administer more gel if necessary

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How do you know that Ceribell is recording EEG and transmitting data to the cloud portal?

A

Both the record and wifi icons are displayed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What should be done to fix an electrode if the indicator is red?

A

Pause the recording if you are in the middle of a recording before you fix the electrode

Rotate the electrode 5 more times while gently pushing the electrode towards the head, and then squeeze the plunger further to dispense any remaining electrode gel

If further rotation does not work, use syringe provided by Ceribell to add more gel to the electrode

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13