Define Seizure Burden
What would 10% seizure burden mean?
% of time the patient is experiencing seizure or seizure like activity over the last 5 minutes.
That means that 10% of the time over the last 5 minutes the patient was experiencing seizure or seizure like activity
What is our sensitivity for alerting for suspected Status Epilepticus?
95% sensitivity
What are the three main objectives for an ICU In-Service
Why do you use Ceribell
How does it work
How is it specifically going to work at Hospital X
System Setup - How should the patient’s head be supported when setting up the Ceribell EEG?
Roll up one large towel or two small towels to support patient’s lower head, to create a space between patient’s head and pillow to prevent pressure from the headband electrodes
System Setup - What should be done with a patient’s hair when setting up the Ceribell EEG?
You should pull long hair behind the patient’s ears and move the hair bangs out of the way
System Setup - What is the correct position for the headband? 3 bullets
Clasp centered
Right above the eyebrows
Right above the ears
Can you use the Ceribell headband on patients with head wounds?
Yes. You can position the headband to avoid the wound if needed, but you should include a note and notify the physician so that the physician is aware of how the headband was placed
What does it mean when the electrodes on the recorder screen are shown in grey?
Cable?****
What is the correct way to prepare the electrodes?
While applying gentle pressure, rotate the electrodes back and forth for 7 complete rotations, and then use two thumbs to push the plunger
I rotated one electrode 7 times and deployed the gel, but the electrode light in the recorder remains red. What should I do?
If you have prepared all of the electrodes and the indicator is still red, rotate the electrodes 5 more times until the light turns green. Use a syringe to administer more gel if necessary
How do you know that Ceribell is recording EEG and transmitting data to the cloud portal?
Both the record and wifi icons are displayed
What should be done to fix an electrode if the indicator is red?
Pause the recording if you are in the middle of a recording before you fix the electrode
Rotate the electrode 5 more times while gently pushing the electrode towards the head, and then squeeze the plunger further to dispense any remaining electrode gel
If further rotation does not work, use syringe provided by Ceribell to add more gel to the electrode