Leakage radiation must be less than _____ mR/hr at a distance of 1m from the housing
100
True or false: A control panel msut have a visible signal indicators only.
False, a control panel msut have a audible/visible signal indicators
The SID indicators must be accurate to within ____ of the indicated SID
2%
True or false: The x-ray beam and light must coincide within 5% of the SID
false
PBL stands for
Positive Beam Limitation
True or false: The image receptor alignment is unnecesarry as long as the light field and x-ray beam are aligned to each other
false
The total filtration for 50 to 70 kVp x-ray machine should be
1.5 mm Al equivalent
The added filtration in a 70 kVP x-ray machine should be
1.0 mm Al equivalent
The variation in x-ray intensity should not exceed
5%
The maximum acceptable variation in linearity is ___% from 1 mA station to an adjacent mA station
10%
True or false: The exposure control in a stationary x-ray unit should be atleast 6 feet or 2 m cord
False
The SSD in stationary fluoroscope should not be less than ___cm
38
The SSD in mobile fluorsocope should not be less than ___ cm
30
The _______ assembly serves as the primary protective barrier in fluoroscopic units
Intensifying screen or Image receptor
The primary barries of fluoroscopic protective units must be ____ Pb equivalent
2 mm