Physics and Monitoring Flashcards
3 ways to minimize oxygen when using anesthesia machine as an ICU ventilator
- Use electrically powered ventilator
- Converting ventilator drive gas to air instead of oxygen for pneumatically driven bellows ventilators
- Using low fresh gas flows
Pneumatically-driven bellows ventilators generate (inspiratory/expiratory) effort. Volume of ventilator drive gas that is roughly equivalent to ______ is injected into the bellows chamber.
inspiratory
the pt’s tidal volume
In most pneumatically-driven bellows ventilators, the ventilator drive gas is _________. But it can be converted to use ____ instead to conserve it.
oxygen
air
The auxiliary oxygen tank will begin to deliver oxygen to an anesthesia machine if the pipeline pressure drops below __ PSI
45
Ultrasound is a high frequency sound with vibrations above _____ cycles per second
20,000
A full E-cylinder contains ___ L of oxygen at ____ PSIG
660L
2000 PSIG
How to calculate time remaining before a tank is depleted
Liters remaining / (current flow rate in Liters/min)
or
take current pressure (PSIG) divided by 3, then dividing by flow rate
When a vaporizer is designed for a volatile anesthetic with a lower saturated vapor pressure (SVP)is misfilled with a volatile anesthetic with a higher SVP, the output volume percentage will be (higher/lower) than the [ ] set with the dial
higher
- the higher the SVP, the higher the partial pressure of anesthetic, and therefore the higher percentage of total gas leaving the chamber
the adjustable pressure limiting (APL) valve, aka the _______ valve, cannot be positioned between the ____ and the _____ to prevent rebreathing of CO2 in a traditional circle system
pop off valve
patient and inspiratory
_______ is the most common breathing system in the US
The circle system
A unidirectional valve must be positioned between the ______ and _____ in order to prevent rebreathing of CO2 in a traditional circle system
patient
both the inspiratory and expiratory limbs
Fresh gas flow cannot come into the circle system between the ____ and the ____ in order to prevent rebreathing of CO2 in a traditional circle system
patient and expiratory valve
The circle system can be used in 3 different ways depending on the fresh gas flow
- semi-open system
- semi-closed system
- Closed system
Semi-open system
high fresh gas flows with no rebreathing
Semi-closed system
lower fresh gas flows with some rebreathing
Closed system
fresh gas flow matches pt consumption with complete rebreathing
_______ protect persons from electrocution by turning a normal “grounded system” (that exists outside the OR) which only needs a single fault to cause electrocution into a “protected” system in which two faults are needed to deliver a shock
Line isolation systems
- Isolation transformer + line isolation monitor
Lasers in the OR is assoc with (4)
- gas embolism
- retinal injury
- OR fires
- laser plume creation
*NOT shown to cause defibrillator discharge
Wandering baseline ECG tracing, especially in leads III and aVR, is commonly caused by _____
loose ECG electrode
Poor skin contact
moving pt (respirations)
Lead II measures the voltage between the ___ and ___ electrodes
Red (left leg)
white (right arm)
Lead III measures the voltage between the ___ and ___ electrodes
red (left leg)
black (left arm)
Lead aVR (augmented vector right) measures the voltage between the ___ and ___ electrodes
white (right arm)
combination of black (left arm) and red (left leg)
Color code these gas types: Carbon dioxide - Nitrous oxide - Nitrogen - Helium - Oxygen - Air -
Carbon dioxide - gray
Nitrous oxide - blue
Nitrogen - black
Helium - brown
Oxygen - green
Air - yellow
The doppler effect can be used in ECHO to determine both the ____ and ____
direction
speed of flow
_____ technique is the most commonly utilized invasive method for assessing cardiac output
Thermodilution cardiac output
Induction agent that most enhances motor evoked potential amplitudes
Ketamine
Somatosensory evoked potentials (SSEP) provides information regarding the integrity of the _____ and are measured by stimulating the peripheral or cranial nerve and measuring the resultant EEG
ascending neurons
Motor evoked potentials (MEP) are used to evaluate the integrity of the ______.
descending motor neuronal pathways
Etomidate may enhance ____ evoked potentials but does not increase ____ evoked potentials
somatosensory evoked potential
motor evoked potential
a 5 MHz probe will provide (better/worse) tissue penetration than a 7.5 MHz probe
better
Higher frequency (lower wavelength) means better ______ but worse _____
resolution
penetration (depth)
Lower frequency (increased wavelength) gives better _____ but sacrifices _____
penetration (depth)
resolution
Velocity =
frequency * wavelength
VWF
In a pt with Fontan circulation, ______ circulation bypasses the heart to reach the lungs. Used for pts with ______ cardiac system.
venous univentricular - tricuspid atresia - hypoplastic L or R heart - double inlet L ventricle
In Fontan circulation, ______ pacing leads may be placed for permanent pacing rather than _____ pacing, d/t access.
Epicardial
Transvenous
A curved array US probe maximizes _____, and is better at _____.
returning US waves
deeper structures
A Linear array US probe maximizes _____, at the expense of _____
resolution
less penetration
Cerebral perfusion pressure =
Spinal cord perfusion pressure =
MAP - ICP
MAP - CSF pressure
*Two most common way to improve SCPP during aortic surgery is either increase MAP or decrease CSF pressure using lumbar CSF drain
SVR = ?
80 * (MAP - RAP) / CO
R atrial pressure = CVP
Half tank of N2O would read a pressure of ___
745 psig
- regardless of the volume of N2O in the tank, the pressure will read 745 psig until ALL liquified gas is used up (~250 L or 16%)
A full tank of N2O contains ____ at a pressure of 745 psig
1590L
atmospheric pressure at sea level is ____
14.7 psig
A full and Half tank of O2 would read a pressure of ___
Full: 2200 psig
Half: 1100 psig
A full tank of O2 contains ____ at a pressure of 2200 psig
660L
Raising or lowering a pressure transducer (ie. art line, pulm a. cath, intracranial pressure) relative to the pt will alter the pressure reading.
a 10 cm change in height will alter pressure reading by _____ mmHg
- 5mmHg
- Raising transducer (lowering pt): lowers the pressure reading
- Lowering transducer (raising pt): increases the pressure
What happens to the pressure reading if you:
- Raising transducer (lowering pt):
- Lowering transducer (raising pt):
- Raising transducer (lowering pt): lowers the pressure reading
- Lowering transducer (raising pt): increases the pressure
- 10 cm change in height will alter pressure reading by 7.5mmHg
The bellows within the circle system is in a (high/low) pressure system, and will STILL function in the presence of leaks in the low pressure system
High
The flowmeters and vaporizers within the circle system is in a (high/low) pressure system, and will NOT still function in the presence of leaks in the low pressure system.
low
- could lead to administration of hypoxic gas mixtures or insufficient amts of anesthetics
The adjustable pressure limiting (APL) valve is typically taken out of the circuit when the ventilator is switched to mechanical ventilation mode. What does this mean?
The system will function the same regardless of whether the APL valve is open or closed
Which one consumes more gas, using gas from the pipeline supply or E-cylinders?
E-cylinders
- so you should hand ventilate if there is a failure with the pipeline supply
The a typical ascending (standing) bellows ventilator, the bellows (rise/fall) during pt exhalation
rise
*the bellows can act as an indicator of a leak or disconnected circuit
The a typical ascending (standing) bellows ventilator, the bellows (rise/fall) during pt inhalation
fall
*the bellows can act as an indicator of a leak or disconnected circuit
3 most important preop anesthesia machine checks
- Oxygen analyzer calibration
- low pressure circuit leak
- circle system test
A suction bulb is attached to the _____ and must remain collapsed for at least _____ seconds to pass a negative pressure leak test. If this is successful, the flowmeter or vaporizer is considered safe.
Common gas outlet
10 s
Fresh gas flow must be greater than or equal to ___ times minute ventilation in Mapelson ____ circuits.
1-2x
D, E, F
Mapelson D, E and F are ____ circuits that are not very efficient for spontaneous ventilation, but is most efficient for ______.
Semi-open circuits
Controlled ventilation
Advantages of a Bain circuit (Mapelson D) (4)
- conservation of moisture d/t partial rebreathing
- warming of fresh gas inflow
- scavenging waste is easier from APL valve
- advantageous when access to the pt is limited
Which mapelson circuit requires higher fresh gas flows (2.5x minute ventilation)?
D (bain)
Mapleson __ is better for spontaneous respirations, and ___ is better for controlled ventilation
A - spontAneous
D - controlleD
If a pt is able to do a head lift, or hand grip, or effective cough, the TOF ratio is ____
0.7 < TOF Ratio < 0.9
T4:T1 height ratio
How much receptor is suppressed? 1 Twitch 2 Twitch 3 Twitch 4 Twitch
1 Twitch: > 90%
2 Twitch: 80-90
3 Twitch: 70-80%
4 Twitch: 65-75%