Physics And Instrumentation Flashcards

1
Q

Amplitude

A

Indicates strength of sound wave
Measured in dB
measured as the difference between the peak pressure in the medium and the average pressure.
Adjusted by changing transmit power

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Wavelength

A

Distance between two successive waves

Measured in length (m\mm)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Frequency

A

Number of wave cycles (oscillations) per second
Measured in Hz
100 oscillations per sec = 100 Hz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Audible sound

A

20 - 20000 Hz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Infrasound

A

<20 Hz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Ultrasound

A

> 20KHz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Frequency of Ultrasound for echo

A

1.5 - 7 MHz

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Propagation Velocity

A

speed at which the wave propagates through the medium.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Propagation velocity of various tissues

A
Lowest to highest
Air
Fat
Soft tissue (I.e heart)
Blood
Muscle 
Bone
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Propagation velocity equation

A

Propagation velocity =frequency x wavelength

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Relationship between frequency and resolution and penetration

A

Higher frequency = better resolution = poorer penetration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Acoustic impedance

A

Resistance to ultrasound transmission

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Two types of reflection

A

Speculator (mirror-like) reflection

Backscatter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Specular reflection

A

‘Mirror-like’
occurs at tissue boundaries where the reflector is relatively large (at least two wavelengths in diameter) and smooth
To maximise energy reflected, incoming beam should be perpendicular to the reflector as possible

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Backscatter reflection

A

occurs with small and/or rough-surfaced structures, where the reflected ultrasound will scatter in many different directions. the returning signal will be weaker than from a specular reflector, but will not be dependent on the angle of the incident (incoming) ultrasound beam.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Rayleigh scatterers

A

Scatter equal in all directions (e.g RBCs)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Attenuation

A

As an ultrasound pulse travels through tissue, it will gradually lose energy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Define HID

A

depth (in cm) in soft tissue in which the intensity of the ultrasound is reduced by 50 per cent, and depends upon the frequency (f) of the ultrasound emitted by the transducer, measured in MHz.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

HID Equation

A

HID (soft tissue) = 6/f

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Refraction

A

change in direction of an ultrasound pulse as it passes across a boundary between two tissues (or materials) of different acoustic impedance.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

What is a transducer?

A

Probe that generates ultrasound and works as transmitter and receiver of ultrasound waves

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Piezoelectric effect

A

Voltage applied –> piezoelectric crystals oscillate –> change shape –> generate ultrasound

Ultrasound sound wave returns –> crystals oscillate and change shape –>creates electrical voltage later used for image analysis.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Phased array transducer

A

consist of an array of piezoelectric elements (typically 128 for a 2D echo probe, several thousand for a 3D probe).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Structure of an ultrasound transducer

A
Cable 
Backing layer
Piezoelectric elements 
Matching layer
Acoustic lens
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

Backing Layer

A

Has high impedance and is designed to absorb ultrasound and ‘damp down’ reverberation (‘ringing’) of the piezoelectric elements.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

Matching layer

A

improves the impedance matching between the elements and the body.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

Near field

‘Fresnel zone’

A

US beam remains cylindrical for a short distance after leaving transducer
Imaging quality best within the near field and maximising depth of near field is important for image optimisation
length of the near field is greater at higher transducer frequencies and wider transducer diameters.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

Far field

Fraunhofer zone

A

Ultrasound diverges after near field

29
Q

Acoustic lens

A

Helps to focus ultrasound beam

30
Q

Fundamental frequency

A

Original frequency of transmitted signal from transducer

31
Q

Harmonics

A

Multiples of original frequency of signal returning to transducer

32
Q

Second harmonic imaging

A

filters the returning signal to remove the fundamental frequency and build up an image using the second harmonic components of the signal. In so doing, the image resolution improves (because of the higher frequency), particularly for far field structures.

33
Q

Disadvantages of second harmonic imaging

A

requires a higher power output, and it does slightly alter the appearance of myocardial texture and also the apparent thickness of structures such as valve leaflets compared with fundamental imaging.

34
Q

A-mode

A

Amplitude mode
plotted the amplitude of the reflected ultrasound (as a ‘spike’ with a certain amplitude) versus the distance of the reflected signal from the transducer.

35
Q

B-mode

A

Brightness mode

represented the amplitude of the returning signal by the brightness of a dot.

36
Q

M-mode

A

Motion mode
records motion along a single ‘line of sight’, selected by careful positioning of the on-screen cursor across a region of interest

37
Q

2D imaging

A

ultrasound probe sweeps a beam across the heart around 20–30 times per second, creating a series of scan lines (usually around 120) each time it makes a sweep, in order to build up a 2D image

38
Q

Frame rate

A

Number of image frames generated each second

reducing the sector width and/or depth will reduce the time taken to generate an image frame, increasing the frame rate

39
Q

Transmit Power

A

controls the amount of ultrasound energy delivered to the patient, and to minimize the risk of adverse mechanical or thermal effects it is important to use the lowest setting possible

40
Q

Gain

A

refers to the amplification of the received signal to increase the brightness of the displayed images. Gain can be adjusted for the whole image (overall gain) or for part of the image (see time-gain compensation (tGC) below). While a high gain setting can be useful for detecting weaker signals that might otherwise not be visible, it reduces lateral resolution and also increases noise.

41
Q

Time gain compensation

A

known as depth compensation, and corrects for the attenuation of the ultrasound signal that occurs with increasing distance from the transducer. tGC boosts the gain of the signals returning from the far field to ensure an even ‘echo brightness’ across the whole depth of the image.
Slider bars

43
Q

Depth

A

determines how far the ultrasound beam ‘looks’ into the patient and is an important determinant of frame rate. the greater the depth setting, the longer the transducer will have to wait for the ultrasound pulse to make its round trip before repeating the pulse, and so the lower the frame rate.

44
Q

Sector width

A

determines the field of view across which the ultrasound beam sweeps.

45
Q

Grey scale compensation

A

Dynamic range
Adjusts number of shades of grey displayed in image
Chooses degree of contrast

46
Q

Resolution

A

Ability to discriminate between 2 objects that are close together in space (spatial resolution) or two events that occur close together in time (temporal resolution)

47
Q

Spatial resolution

A

2 components :
Axial resolution
Lateral resolution = azimuthal resolution

48
Q

Axial resolution

A

Relates to objects along axis of us beam
Determined by transducer frequency and pulse length
Higher frequency + shorter pulse length = better axial resolution
Typical axial resolution = 3mm

49
Q

Lateral resolution

A

Azimuthal resolution
Relates to objects perpendicular to us beam
Varies according to sector width/depth/gain
Higher gain = worse lateral resolution

50
Q

Temporal resolution

A

=frame rate
Frame rate depends on time taken to collect data required to take one image
M mode higher frame rate compared to 2d imaging

51
Q

Imaging artefacts

A

‘Structures’ and/or distortions seen on echo that are not actually present in the heart or in that location

52
Q

Acoustic shadowing

A

Highly echo reflective structure (e.g. Mechanical valve prosthesis) blocks us from penetrating further causing echo dropout in far field

53
Q

Reverberation

A

Sound encounters 2 highly reflective layers. Sound is bounced back and forth between two layers before travelling back therefore takes longer and presumed to be farther and display additional ‘reverberated’ images in a deeper tissue layer.
Recognised to be artefact as moves in tandem with original structure.

54
Q

Comet tail

A

Similar to reverberation
Produced by front and back of strong reflector
Reverberation space is very narrow and blends into a small band to look like a comet tail

55
Q

Ring down artefact

A

Resonance phenomenon from collection of gas bubbles

Appears similar to comet tail

56
Q

Beam width artefact

A

When us beam is wider than diameter of refractor being scanned. Therefore normal tissues that lie adjacent to lesion are included in the beam width and their echo texture is averaged to be that of the lesion.
Can be reduced by focusing us beam to minimise its diameter.

57
Q

Side lobe artefact

A

Arises from unwanted/unavoidable side beams. Therefore signals returning from side lobe beams are interpreted as coming from main beam, resulting in display of blurred structure some distance away from original

58
Q

Ultrasound processing method

A

returning echo signal at the transducer undergoes a series of initial processing steps which include amplification, tGC and filtering. the video signal is then sent to a scan converter, which converts the signal into a ‘rectangular’ format suitable for display. the resulting data undergo further processing (‘post-processing’) and can then be stored in a digital format and/or can undergo digital-to-analogue conversion to create a video signal for display on a monitor (and/or archiving onto videotape). this process occurs so rapidly that the acquired data can be displayed on a monitor almost in ‘real time’.

59
Q

Unit of measurement of intensity of exposure

A

Watts/cm2
Power per unit area
Maximum intensity within the ultrasound beam

60
Q

Spatial peak temporal average (SPTA)

A

Spatial peak averaged over the duration of exposure (temporal average)

SPTA = Spatial peak avg / temporal avg

61
Q

Biological effects of ultrasound

A
Thermal effects (heating)
Mechanical effects
62
Q

Thermal effects of ultrasound

A
Most notable in TOE
caused by conversion of mechanical energy of us into heat energy as it passes through the tissues 
Amount of heating is related to:
Transducer frequency 
Transmit power
Focus 
Depth
63
Q

Mechanical effects

A

Mechanical effects of ultrasound can also be measured by mechanical index (MI), which is the peak negative (rarefactional) pressure divided by the square root of the transducer frequency. An MI of <1 is considered safe.

include cavitation, in which gas bubbles are created as ultrasound passes through the tissues.
When bubble contrast is used it can cause resonance and bubble disruption

64
Q

Focus

A

Fine tuned with phased array transducers

Concentrates on area of interest

65
Q

Speed of sound in tissue

A

1540m/s

66
Q

Equation for propagation speed

A

Propagation speed = frequency x wavelength

67
Q

Period

A

Time taken for one cycle or wavelength to occur

68
Q

Period equation

A

Period = 1/frequency

69
Q

Basis of continuity equation

A

Law of conservation of mass

70
Q

EROA equation

A

EROA = regurgitant flow rate / flow velocity