PET Flashcards
3 steps to PET
Administer tracer
Image acquisition
Image reconstruction
PET tracers
18F-FDG
18F-DOPA
11C-choline
11C-methionine
11C-acetate
68Ga-DATA-NOC
Comparison to Gamma camera
Full ring of detectors rather than floating heads
No collimator - uses coincidence
Rarely stand alone - often compatible with CT or MR
Often use step and shoot approach - have number of bed positions and time per bed position and overlap (although system exists that uses continuous bed motion)
LOR
Line of response - determines relationship between annihilation location and detection. PET more sensitive because collimators usually absorb a lot of emissions
Types of event
True coincidence
Scattered coincidence
Random coincidence
Attenuation
Block detectors
Scanners have multiple rings
Detector rings typically have many blocks.
Scintillator crystal attached to 4 PMTs
Cuts in crystal act as light guides
Interaction position determined by ratio of signal in each PMT
System could have 10000-20000 crystals 4-6mm in size with only 1000 PMTs
What do we want from a scintillator?
Same as gamma camera - but at higher energy
(high efficiency at stopping, high probability of PE not CS, high conversion of energy to photons, transparent to own emissions, good matching between light and PMT response, mechanically robust, short scintillation time is more important because of higher count rate in PET)
What scintillator do we use?
Now LSO and LYSO (cerium doped).
NaI was used but wasn’t ideal, then BGO due to better stopping efficiency and PE interaction rate.
Signal processing - what do we need
Two signals in coincidence - both signals in energy window (which is usually worse than SPECT) gives line or response.
Coincidence window
Allowable time window between the two interactions for them to be considered a pair. Only paired events used in data.
Amplitude discriminator
Amplitude = energy. Energy resolution in PET worse than gamma cameras 350keV to 650keV. Further scatter correction necessary.
Corrections in PET
Randoms correction
Scatter correction
Normalisation
Dead time correction
Decay correction
PET scanner calibration
Sinogram
Events categorized by plotting each LOR as function of angular orientation vs displacement from centre of gantry.
List mode also available
2D vs 3D
3D has no septa, better sensitivity, more sensitive to quality of scatter correction.
2D has septa, and planes are processed independently by 2D algorithm.
Bed position overlap
Sensitivity will vary through the scanner depending on the overlap, no overlap sensitivity will vary from 100-0 but a lot of overlap sensitivity will be more steady - takes longer.