Performance standards, Analyser & Assays Flashcards
What are “allowable limits of performance”?
It is the normal range for each component measured. e.g. +/- 3.0mmol/L for Cl means w/in +/- 3mmol/L from true value
The allowable limits of performance for sodium is +/- 3.0 mmol/L. If a lab can measure plasma sodium with a SD of 3.0 mmol/L, is that satisfactory? Explain
your answer
No bc if 1 SD is 3mmol/L that means the range is 6mmol/L in total. To determine the required mmol/L of 1SD (so it fits the limit of +/- 3mmol/L) must calculate [allowable limit/4]. */by 4 bc thats the total SD range if SD were to be 2SD (=cover 95% results) ==> 3mmol/L / 4 => 0.75mmol/L (1SD)
What is a proficiency program, when is it necessary and of what use is it?
- External Quality Assurance program
- Long-term QC that measures the the lab’s performance every month, & compares results with other labs
- compares analytes/ machine/ method used
Compare the way we use internal and external QC programs.
EQA: Compares results of a lab w/ other labs - are using the same instrument / reagent. Done every month
Int. QC: monitors precision & accuracy of assay & analyser. Done daily
What type of auto-analyzer would be required to measure a steroid present at a concentration of 12 pmol/L. In addition, what type of assay principle would be required, and what would the calibration curve look like? *
Sml [ ] require more sensitive method => Immunoassay, competitive assay (=inversely proportional relationship)
How do titrimetric assays differ from competitive assays? *
Titrimetric: sample put first then incubated then add labelled Ag
Competitive: Sample Ag & labelled Ag added at same time
Compare the different type of solid phases used by Roche Elecsys and Siemens Immulite assays.
Roche: Streptavidin coated magnetized beads
Siemens: Capture Aby coated on bead
What does the term homogenous signify, and what advantage do these have compared to heterogenous methods
Homogenous: all in liquid state. Everything is added in reaction = easy to automate & faster (less sensitive)
Heterogenous: Solid & liquid state. Removal/wash steps = time consuming but more sensitive
The Roche Elecsys system is subjects to two sources of interference that are not seen in most other sensitive immunometric assays. What are these?
- patients taking large doses of biotin (vit B7) = interfere
- Anti-ruthenium Aby in patient samples = false positives
What are the two major auto-analyzer types in a clinical biochemistry laboratory and give an example of major differences in the analytical principles of the chemistries they employ
- General chemistry analysers: Larger MW e.g. proteins, smaller vol (paeditrician). Homogenous assay
- Immumoassay analysers: Smaller MW e.g. hormones. Heterogenous assay (specific)
What is chemiluminescence, and how does it differ from fluorescence or light colorimetry, and why do immunoassays using chemiluminescence yield a higher sensitivity
- no light source bc a photon light emitted from reaction = higher sensitivity from high signal
- light emitted is being measured
What is a dry chemistry analyzer?
- slides with layers of reagents (not use liquids)
Difference b/w chemiluminescent assay and chemiluminometric assay
Chemiluminescent:
Chemiluminometric: use chemiluminescence report system
Describe 2 errors in immunoassays (& suggest solutions if possible)
- hook effect: Excess Ag saturates capture & labelled Aby. Solution: 2-step assays
- interfering Aby from pts sample: pts Aby can cross link
& block sites for Ag = false pos signals