Patient dose calculation systems Flashcards
1
Q
Why do we do independent dose calculations? (3 marks)
A
- legally required to do so
- checking the TPS is correct
- can be factor based or beam model based
2
Q
How are factor based calculations carried out? (4 marks)
A
- assume the patient is a homogeneous rectangle
- identify the differences from reference conditions
- correct for those differences
- can use the ESTRO formulism
3
Q
Explain the ESTRO formulism for isocentric treatments? (5 marks)
A
D(z,c) = D_R x U x O(c) x V(z,c)/V(z_ref, c_ref) x T(z,c)
where: D_R = dose per MU under ref conds U = MUs O(c) = head scatter correction V = scatter volume ratio TPR = distance correction
4
Q
What are the two main issues and solutions with using the ESTRO formalism? (4 marks)
A
- volume scatter ratio is not easily defined, use phantom scatter ratio instead
- beam size is defined at the surface, use isocentric method and define beam size at isocentre
5
Q
What is the practical version of the ESTRO formalism? (5 marks)
A
D(z,c) = D_R x U x S_p(z,c) x S_c(z,c) x TPR(z,c)