Pacemakers (edit) Flashcards
In DDDR mode explain tracking of intrinsic rate Vs sensor indicated rate
Intrinsic atrial - AsVp
Sensor indicated - ApVp
Note* Whatever is faster is in control
Purpose of Upper tracking rate and Upper sensor rate
Rate limits for atrial tracking that are programmable.
Prevents tracking of atrial tachycardias
True or false
In DDI mode An atrial event INSIDE the PVARP inhibits scheduled atrial stimulus but will not start AV interval
False
An atrial event sensed OUTSIDE the PVARP will INHIBIT a scheduled Ap (I.e. sensed) but will not start a AV interval
Note* basically As doesn’t not start AV delays is DDI mode
At what rate do Vp events occur following As events during DDIR mode pacing?
At the sensor indicated rate
Note* Vp after As is at sensor indicated rate not base rate (paced at base rate in DDI mode only)
Pacing mode with continuous Ap regardless of intrinsic atrial activity
DVI/R
Note* sensing only occurs in the ventricles
True or false
Programming rate response on in a chrontropically competent patient is beneficial
False
It would unnecessarily increase Ap with possible competition between the intrinsic and pacing rates
- in the long run this would be unnecessary battery depletion
True or False
Triggered mode pacing occurs at the LRL
True
But a As/Vs event also triggers an immediate pacing output (rather than inhibiting)
Blanking periods in ODO, OAO and OVO are…
automatically minimised to maximise sensing windows
Note* marker channel May display FFRWOS that otherwise would not appear if blanked as usual
Decremental Device testing in clinic unmasks twitching at high pacing rates. What should be done?
Test twitching from high to low voltage and see where it stops. There should be a 2V safety margin if possible.
CRT post implant optimisation includes
- Checking morphology is narrow with optimal VV and AV delays
- Checking for change in morphology during threshold testing (confirms true RV and LV capture)
- ECHO optimisation of AV delays if necessary
Identify the appropriate steps for device check
- Read all previous reports
- Take snapshot presenting rhythm
- VHR/AHR episodes
- Sensing test - underlying rhythm
- Threshold test - look for change in morphology/EGM to confirm capture
- Parameters - adjust as appropriate according to testing results
Clinical significance of pseudofusion
Battery depletion
*occurs when intrinsic heart rate is very close to the lower rate of the pacemaker - Pseudofusion is not dangerous to the patient.
What is important to look for during device threshold checks?
Loss of capture
Change in EGM/ ECG morphology to confirm capture.
*no change but pacing spike = pseudofusion or true pacing
* Change and pacing = fusion or true capture
In what inherited arrhythmia can a dual chamber ICD be indicated?
LQTS where AP is required to maintain regular rhythm and prevent long short intervals
extra lead for discrimination is not indicated
Bipolar sensing WITH unipolar pacing suggests…?
Issue with lead impedance/integrity or high thresholds
Unipolar sensing WITH Bipolar pacing suggests…?
Noise on Bipolar channel unable to be blanked by changing sensitivity settings
Are epicardial device MRI compatible?
No!
*Epicardial leads have not been MRI tested and thus not licensed to have MRI performed with patients implanted with such devices.
How is it determined if a signal is seen by a/the device?
Marker channel annotations indicate sensed and paced events as well as detection and therapies delivered
How does FFRWOS occur
Tip of atrial lead is close enough to ventricle to sense ventricular depolarisation
*i.e. if in right atrial appendage which sits right over RV then oversensing can occur
Surgical Vs congenital escape/Junctional rhythm - which is more reliable?
A surgical escape rhythm is less reliable than a congenital escape rhythm
Noise always results from…?
A
Conductor fracture or insulation breach
EMI
TENS
*noise on a lead is not normal
What do conductor and insulation failure affect?
Thresholds (trending or sharp increase)
Impedances
*these done always change dramatically particularly if there is a micro fracture/micro dislodgement
*Occasionally, pacing thresholds and impedance rise, typically gradually over months, without any detectable fracture from development of scar tissue at electrode myocardial interface (exit block) or the deposition of calcium hydroxyapatite crystals at the lead-tissue interface
What is exit block?
development of scar tissue at the electrode myocardial interface (exit block)
How has exit block been resolved?
Steroid elution in current-generation pacing leads has virtually eliminated the risk of exit block
Predictors of good lead fixation
EGM widening and ST segment elevation and good sensing