pacemakers Flashcards

1
Q

what are the two ways to correct symptomatic bradycardia?

A

atropine and or pacemaker

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

when are temporary pacemakers used?

A

primary in emergency circumstances when the need for cardiac pacing is brief or self limited.

immediate correction of symp brady during severe sinus node dysfunction or advanced AV block

prophylactic intervention prior to procedures requiring a general anesthetic in patients who have underlying bifasciular block

tachyarrythmias that can’t be suppressed with drug therapies and the temporary pacer would be used to overdrive or break the rhythm

part of diagnostic procedures such as electrophysiology studies to evaluate the conduction system and the potential for arrythmias

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

what are the two types of temporary pacing?

A

transvenous and epicardial pacing and transcutaneous pacing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

what is transvenous pacing? can u move around?

A

a temp pacing catheter placed thru either the femoral, brachial, jugular, or subclavian veins to stimulate the endocardial surface of the right ventricle. no bedrest cause it can move it

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

what is epicardial pacing?

A

epicardial wires are temporarily sutured onto the epicardial surface of the RA and/or RV during cardiac surgery. the wires can exit through the chest wall and a re taped to the chest wall for quick access.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

what is trancutaneous pacing?

A

external pacing pads located on the anterior or posterior chest wall for emergency thranathoracic pacing with the use of the pacing mode on the LifePak defibrillator

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

what is the most common form of temporary cardiac pacing?

A

ventricular demand pacing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

demand pacing utilizes what?

A

a ventricular lead which sense and pass and when the lead senses the patients QRS the pacing stimulus is inhibited and in the absence of pt QRS it will deliver a stimulus

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

what does the pacing stimulus generate on an ECG

A

an electrical artifact referred to as a pacing spike

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

what is pacemaker syndrome?

A

symp that patients may experienced when ventricularly paced for a long period of time. symp are fatigue, dizziness, syncope and pulmonary congestion. results from loss of AV synchrony and VV synchrony

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

what does this loss of AV ad VV synchrony result in?

A

dec CO d/t loss of atrial contribution and ventricular filling

atrium contracting against a closed alive causing backflow of blood into pulmonary veins which can trigger atrial baroreceptors and cause hypotension

ventricular pacing is the only option for patients who have undergone AV node ablation for afib

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

what is temporary atrial pacing? where is the lead? what node needs to function here? what is this useful for?

A

single lead placed in right atrium. once chambers are activated the electrical impulse is conducted through the normal conduction pathway to stimulate the ventricles. requires a healthy AV node and is useful for…

improving cardiac output by providing synchronized contraction of the atria and the ventricles and
-overdriving or suppressing supra ventricular tachydysthymias

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

what is atrioventricular sequential pacing or dual chamber pacing?

A

pacing lead n both r atrium and r ventricle and requires an intact. requires orderly and sequential contraction of the atria and ventricles. AV synchrony provides the atrial contribution of atrial kick to ventricular filling and sig adds to CO. this atrial contibtribution is estimated to be 20-30% of the cardiac output. in its with marginal cardiac reserve acute MI pt (ex), the loss of AV synchrony may result in acute CHF and other decompensatory changes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what is BiVentricular pacing or cardiac resynchronization therapy (CRT)?

A

lead in both ventricles. accomplished w a transenous lead placed in the RV (endocardial placement) and a second ventricular lead is threaded through the coronary sinus to be placed in a vein located on the epicardial surface of the left ventricle. the concurrent stimulation of the right and left ventricles results in a synchronous myocardial contraction reducing heart failure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

what is an ICD?

A

implantable cardioevrter defibrillator.

it is placed i pt with inc risk of sudden cardiac death due to lethal cardiac arrytmias. usually post MI and have congenital prolonged QT intervals and will keep arresting.

atrial and ventricular leads are located in the RA and RV. programmed to overdrive pace, cardiovert or debrillate VT or Vfib

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

what are the components of pacemakers?

A

a temp consists of a transvenous bipolar catheter (lead) which is attached to an external pulse generator .

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

what do bipolar pacemaker leads have?

A

2 electrodes in contact with the cardiac tissue. the distal or stimulating electrode is usually neg (cathode) one and the positive (anode) is usually 1cm proximal to the tip

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

what do unipolar pacemaker leads have?

A

infer sued has one neg electrode in contact w heart and a positive extra cardiac electrode which is part of the pulse generator. the unipolar gives large ECG spikes easily recognizable bit is susceptible to interference from large groups of muscles (eg. pectorals)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

which pacemaker lead is more common?

A

bipolar bc generates a small EC spike but not susceptible to outside muscle interference

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

what is the pulse generator?

A

includes the power source (or battery) and the electronic circuitry responsible for sending out appropriate timed signals or creating an electrical pulse and for sensing cardiac activity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

what is the catheter lead system?

A

connects the pulse genitor to the heart. it includes and insulted conductive wire

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

where is the pacing catheter in relation to the heart?

A

it is in direct contact with the heart. usually the septal wall of the right ventricle. the pacing catheter delivers the electrical impulse to the myocardial and sense intrinsic cardiac activity. sensed intrinsic cardiac activity may inhibit or generate a pacing imupilse

23
Q

what is sensitivity?

A

define as the pulse generators ability to sense or see the persons own activity. dial of .8-20 mv. numbers represent amplitude of the QRS the pacemaker is ‘seeing’ so lowest is most sensitive. referred to as demand pacing

24
Q

(sensitivity) at the highest setting the pace will not sense intrinsic activity. this is referred to as?

A

asynchronous or fixed rate

25
Q

what is output/milliamps?

A

electrical activity delivered from the pulse generator through the pacing electrode to the myocardium to initiate depolarization. the amount of current delivered is measured in milliamps. the dial not he temp pacemaker inc in number as the output is increased.

26
Q

what is capture?

A

when sufficient electrical current or output is delivered to the heart muscle, depolarization occurs; this is known as capture”

27
Q

depolarization can also be caused by what? (other than the pulse generator) what is the name for this

A

by intrinsic cardiac activity occurring simultaneously w the pacemaker producing a complex called a fusion beat

28
Q

what does a fusion beat look like? when does it usually occur? is this dangerous?

A

has features of both the intrinsic and artificially paced morphology. can vary in shape or morphology depending upon which stimulus contributed most to depolarization

when pacemaker is close to own pt heart rate. not dangerous

29
Q

what is threshold?

A

the amount of elecrtrcal current required to cause depolization is called the threshold. determined by

-inc the pacing rate over the patients intrinsic rate so that paced rhythm can be identified or observed
-dec the MA or output while the paced rhythm is being observed to a point where the paced rhythm is no longer seen
the MA is inc again to the point that the pacemaker captures the ventricles and a paced rhythm is observed not he monitor. the point that pacing returns is known as the threshold

30
Q

what is encodcardial tissue fibrosis and what will this cause in re: pacemaker?

A

inflm response caused by the constant stimulation of the pacing catheter
irritated cardiac muscle cells have a dec response to electrical stimulation of 2-3 days which affects threshold and makes it ned for more electrical current (inc MA) to be delivered to cause depolarization. therefore, the pacemaker output is usually set 2 1/2 times higher than the threshold

31
Q

what is rate? how is it set?

A

number of electrical stimulation/min. ordered by dr. f pt is pacemaker dep, rate must be adequate to provide good CO. if pt is not dep then the rate is adjusted to a lower limit to provide back up and allow the pt to be in their own rhythm. dec incident of competitive pacing

32
Q

first pulse generators are at a field rate or synchronous which means what? what problem does this cause?

A

meaning electrical signals fire at a pre-set rate without regard for the patients own heart rate. this occasionally created a problem with competitive rhythms causing palpitations in some patients. also created the potential for the pacemaker to fire not he vulnerable period of the t wave and cause a life threatening dysthymia such as vfib. resolved w the development of sensing circuits that made demand (synchronous) pacemakers possible

33
Q

why are pacemakers in the synchronous or demand mode better?

A

they sense the patients intrinsic heart rate which prevents or suppresses the firing of the pacemaker when the intrinsic rate is greater than the set rate of the pacemaker

34
Q

what do temporary AV sequential apcemerks do?

A

they combine synchronous and asynchronous functions and the pacing leads are situated int eh atria and centricle and have the ability to pace both chambers

35
Q

is optimum sensitivity the same in all patients?

A

no it varies. dr must adjust. usually set at max or demand

36
Q

what is the escape interval?

A

it is the interval from the patients last beat to the first paced complex

37
Q

what is the basic rate interval?

A

the interval between paid beats, usually the escape and the basic rate interval are the same

38
Q

what is the fusion beat?

A

paced activity and patient intrinsic activity occurring at the same time

39
Q

what is the pacing spike?

A

artifact produced when the pacemaker fires

40
Q

what is the magnet rate?

A

results from placing a magnet over the pacemaker generator, this ‘turn off’ the sensitivity of the pacemaker causing it t function asynchronously. this is done occasionally n perm pacemekrst o test its pacing function as well as the magnet rate is indicative of battery life

41
Q

what is the magnet use with an ICD?

A

when ICDs are misfiring so when pt is in sinus tach, conscious with a good bp and device is firing. turns it off but does not disable pacing

42
Q

what is undersensing ? what is the soln?

A

pacemaker doesn’t pick up on intrinsic beats and competitively fires. may cause tachyarrythmias due to doubling of pt hear rate. and result in a pacing spike landing in vulnerable period of the t wave triggering V.F

SOln: dr order to inc sensitivity. if this does not sole the problem and the patients has a stable underlying rhythm, dr may order to turn off pacemaker

but if they do not have stable rhythm then may accept failure to sense and ask that it is monitored closely

reposition pacemaker (Dr)

43
Q

what is oversensing? soln?

A

if pacing circuity is too sensitive, it may inhibit from firing due to non cardiac electrical activity (eg. muscle movement) this can cause bradyarrhythmias

obtain dr order to dec sensitivity

44
Q

what is outside interference? soln?

A

pacemaker generators are better shielded than in the past from 60 cycle interference, i.e. from telemetry units, infusion pump or other equipment normally seen on cardiac ward. there is somer risk if OR if cautery machines, i.e. electrosurgery units are in use bc they emit electrical signals that a pacemaker may interpret as ventricular depolarization

soln: in pacemaker dep pt then turn the sensitivity off so that asynchronous or fixed pacing occurs

45
Q

different ways failure to capture will occur? soln for each?

A

catheter tip not in correct w cardiac muscle. the generator will not capture or depolarize the ventricles and this problem may be intermittent as the pacemker catheter floats in and out of touch w the muscle wall. patients with dilated chambers are difficult to position with tip floating out of position (gotta change the pt position and tell dr)

pacemaker is in area of damage and capture not possible. if this area becomes ischemic, cells will have a higher stimulating threshold (dr needs to inc MA)

marginally perfused cells which are exposed to antiarrythmic drugs will have higher thresholds (soln: with dep pt turned to left side and check all connections, or MA inc and battery replaced and if no work prepare for BCLS, ACLS, or another option is to opting he transcutaneous pacer from emerg and tell dr stat

46
Q

what is failure to capture?

A

emitted stimulus does not initiate depolarization

47
Q

what ifs allure to pace? problems? soln?

A

no evidence of pacing spikes- fi the patients intrinsic heart rate is above pacemaker rate and pacemaker is inhibited (this is appropriate so no problem)

but if patients intrinsic rate is below pacemakers setting then check generator make sure its on, check all connections change battery, change pulse generator and cable, if underlying rhythm is unstable be prepare for BCLS, ACLS, obtain transcu pacemaker and tell dr

48
Q

what are the indications for external transcutaneous or transthrroacic pacing?

A

provide v pacing quick to allow time to get temp pacemaker
noninvasive pacing here is good to Inc CO for patients with bradyarrhytmias, systole, conduction disturbances occurring with acute MI and for emerg support during perm pacer failure. also used in situations where pacing is needed on a standby basis for rhythms such as sinus pause, rbadycardias or 2 degree heart block type II

49
Q

method of external transcutaneous or transthoracic pacing?

A

pacing occurs via two large electrode pads placed over the precordium and on the posterior left side of the chest beneath the scapula. inc pulse duration and large electrode surface area provides lower current density requirements for capture.

when transcutaneously pacing, the Egg rhythm of the patient must be monitored independently with the 3 lead cable provided on the defibrillator. if the ECG leads are not attached, the unit paces asynchronously. the monitor screen shows a pacing artifact, followed by a wide complex if ventricular capture occurs

50
Q

what are the modes?

A

non demand, fixed or continuous pacing will occur continuously at selected rates regardless of the patients own intrinsic rhyth

51
Q

what is demand pacing?

A

pacing that sense the intrinsic QRSctivity and inhibits pacing stimulus for the cycle. or goes off when no intrinsic beat

52
Q

what is inhibited?

A

means the pacemaker does not fire when it senses pt rhythm

53
Q

what is triggered?

A

means that once a preset interval expires the pacemaker fires or paces the atria or ventricles