Option I: Medical physics Flashcards
Sate and explain that a change in observed loudness is response of the ear to a change in intensity:
- Different people can describe different intensity sounds as appearing to have the same loudness – the frequency of the sound is an important factor.
- Depends on the amount of energy sound waves bring to a unit area every second.
- Depends on the amplitude of the sound. A more intense sound (louder sound) must have a larger amplitude.
- The relationship between amplitude and intensity is true for all waves.
State and explain that there is a logarithmic response of the ear to intensity:
- Sound intensity levels are measured on the decibel scale (dB).
- The decibel unit is one tenth of a base unit and is called bel (B).
- The human ear can respond to a huge range of different sound intensities.
- The decibel scale is logarithmic.
- The scale compares any given sound intensity with intensity at the threshold of hearing (the weakest sound that a person is able to hear)
Intensity
is the amount of energy a sound wave bings to a unit area every second
Intensity level
the amount of energy transmitted
Attenuation coefficient
a constant that mathematically allows us to calculate the intensity of the X-rays given any thickness of material
Half-value thickness
For a given energy X-rays and given material there will be a certain thickness that reduces the intensity of the X-ray by 50%
Acoustic impedance
the product of the density of a substance and the speed of sound in that substance.
Solve problems involving acoustic impedance
Z=ρc
- Very strong reflections take place when the boundary is between the two substances that have very different acoustic impedance. This can cause some difficulties.
- In order for the ultrasound to enter the body in the first place, there need to be no air gap between the probe and the patient’s skin.
- An air gap would cause almost all of the ultrasound to be reflected straight back. The transmission of ultrasound is achieved by putting a gel or oil between the probe and the skin.
- Very dense objects (such as bones) can cause a strong reflection and multiple images can be created. These need to be recognized and eliminated.
Pulse oximetry
- Red and infrared laser light is shone through a thin part of a patient´s anatomy, e.g. fingertips
- The amount of absorbance of each wavelength will depend on many factors but a changing relative absorbance between the two wavelengths can be used to determine the ratio of blood cells and oxygen and those without oxygen.
- This allows the overall oxygen content of the blood to be determines without having to take a sample of blood.
Endoscopes
- An endoscope provides the ability to look inside the human body without invasive surgery.
- A tube is inserted into the body. Within the tube there is a collection of optical fibers that allows illumination from a outside source to reach the end of the tube and illuminate the region under investigation.
- Reflected light is collected using a lens system and further optical fibres are used allow an image to be viewed.
- Many endoscopes include additional instruments including a laser.
Laser as a scalpel and as a coagulator
- A laser focused on a small region can increase its temperature so high that cuts through tissue like a scalpel.
- The laser heating ensures the site id kept free of germs and thus the risk of infections is reduced.
- Also blood vessels and nerves are automatically sealed off.
- A defocused laser beam can stop bleeding by stimulating the blood to form a clot.
Exposure
the amount of electric charge per unit mass, produced in a body due to ionizing radiation.
Absorbed dose
the amount of radiation energy absorbed per unit mass; units are the gray (1Gy = 1 J kg-1)
Quality factor (relative biological effectiveness)
a dimensionless factor that takes into account the fact that different radiations have different effects even when they deposit the same energy in a body.
Dose equivalent
the product of absorbed dose times the quality factor for the radiation involved. It is measured in Sieverts (1Sv = 1 J kg-1)