ONYX Flashcards
Microcatheters compatible with Onyx
- Apollo
- Marathon
- Echelon
- Rebar
ONYX Indication
Pre-surgical embolization of brain arteriovenous malformations (AVMs)
ONYX LES Material Components
- Embolic Agent: EVOH
- Solvent: DMSO
- Contrast Agent: Micronized tantalum powder
True or False: Onyx delivers in an adhesive manner
FALSE- cohesive* manner
Onyx Kit Components
- 1.5 mL vial of Onyx (18 or 34)
- 1.5mL vial of DMSO
- 3 leur-lock syringes:
- 2 (white) - onyx
Onyx Kit Components
- 1.5 mL vial of Onyx (18 or 34)
- 1.5mL vial of DMSO
- 3 leur-lock syringes:
- 2 (white) - onyx
Onyx Kit Components
- 1.5 mL vial of Onyx (18 or 34)
- 1.5mL vial of DMSO
- 3 leur-lock syringes:
- 2 (white) - onyx
- 1 (yellow) - DMSO
Apollo Onyx Deliver Catheter Specs
ID:. 013”
OD: 2.7F (Proximal); 1.5F (Distal)
Dead Space: .23mL (.20mL with adapter)
Marathon Flow Directed Catheter Specs
ID: .013
O.D: 2.7F (Proximal) ; 1.5F (Distal)
Dead Space: .23mL
Echelon 10 Microcatheter Specs
ID: .017
OD: 2.1F (Proximal) ; 1.7F (Distal)
Dead Space: .34mL
echelon 14 specs
ID: .017”
OD: 2.4F (proximal); 1.9F (distal)
Dead space: .34 mL
rebar 14 microcatheter specs
ID: .017”
OD: 2.4F(proximal), 1.9F (distal)
Dead space: .27mL
Shake Onyx vials for ____ mins at a setting of ____
20 mins; setting of 8