Nuclear medicine II Flashcards
Primary interaction NaI in Tc99 imaging
Photoelectric (High Z material)
Energy resolution for Tc 99 m on Gamma camera
10 % (14%, because the FHWM)
Energy resolution
Photopeak broadening as a percentage of photopeak energy
If energy resolution is 10% (10% FWHM) what is the energy window
20%
What is the sensitivity range of well counter (contamination testing)
nCi
Pulse height analyzer window width
20%
What detectors DOES have gas and DOES NOT have NaI
Dose calibrators (before injection measure). - Provide radioisotope info
What is ACLG used for QA
ACLG is used for dose calibrators
- A: Accuracy annual
- C: Constancy daily
- L: Linearity quaterly
- G: Geometry Installation
Dose calibrators have a high accuracy (injection within 20%)
Medium energy collimators
Ga 67
K 81
In 111
High energy collimators
I-131
Accuracy required for dose injected (NRC comission)
20%
Low energy collimator
Tc99 m and I123
Intrinsic
withou collimator (INdependent)
extrinsic
with collimator (EXtra)