NM QC Flashcards
Radio calibrator QC level A
Daily Physical inspection High voltage / bias check Clock accuracy Background Zero adjustment Constancy
Radio calibrator level B QC
Annual
Stability - repeat measurement
Accuracy - readings of a source with a known activity / traceable calc
6 monthly
Linearity - accross a clinically relevant range clinical
Gamma probe QC level A
Each use Power source check Physical inspection Background check Sensitivity / constancy check with a test source
Gamma probe Level B
6 monthly
Stability
Energy spectrum
(Comissioning - Uniformity)
Sample counter QC daily / monthly / annual
Before use Background check Sensitivity check with a test source (Cs137 - long lived) 6 monthly Stability - short term count precision Annual Linearity - vials of different activity same volume accross clinical range Dead time - high activity source
Gamma camera daily level A
Physical inspection Collimator touch pad checks Energy window check Tc99m / other E used BG count rate check Intrinsic/ extrinsic flood - 4Mcts (tc-99m/Co57)
Gamma camera level B
Monthly COR Intrinsic (40Mcts) Extrinsic flood (40Mcts) JAZCZAK - Tomo uni/contrast / Res / atten Corr Uni rods spheres 3m Intrinsic other nuclides. 6m Spatial res / linearity - bar phantom. Multi window spatial resolution Pixel size - abs pixel size Annual Whole body uniformity - spatial resolution in air As req Uniformity calibration - update correction map
PET QC level A
Physical
Point source check
Manufacturers CT checks
PET QC level B
Post maintenance
Uniformity 68Ge cylinder or F18 phantom
Variable.
Normalisation - set response to act within FoV
Calibration - find calibration factor from image voxel intensity to find the true activity concerntration
Annual
Spatial resolution - point source sinogram
Count rate performance - measure count rate of decaying activity
Sensitivity - volume resp of sys to a known activity
I’m quality - NEMA NU 2 phantom hot/cold spheres
CT monthly
CT number check
Pet/CT alignment check
Annual
As per rad phys advise