MRI QA Flashcards
1
Q
What are the ideal requirements for Radiotherapy planning?
A
Flat top couch - no carbon fibre or metal
Lasers - MR compatible
Large bore
Coil supports - preserves skin contour
Location mechanism needs to be compatible with immobilisation devices
2
Q
What considerations need to be made for Radiotherapy QC?
A
Geometric distortion - non linear in static and gradient fields
Effect of RT couch on image quality
Consistency of image quality parameters
3
Q
What are the available QC tests?
A
Magnetic field homogeneity evaluation Slice position and thickness accuracy RF coil checks - SNR, image intensity uniformity Table accuracy Laser set-up Laser movement accuracy High contrast spatial resolution Artefact evaluation Centre frequency Transmitter gain or attenuation Low contrast detectability
4
Q
Why is using the ACR phantom advantageous?
A
Tolerances are documented and there is an accreditation program