MQSA Flashcards
months of mammo needed
3
recall rate
<10%
lay report deadline to patient
within 30 days
patients can obtain ORIGINAL mammos when needed
patients can obtain ORIGINAL mammos when needed
who’s in charge of QC program
Interpreting physician
Required resolution (line pairs per mm)
13 in anode to cathode direction
11 lp/mm in left to right
phantom fiber etc
4 fibers
3 microcalc clusters
3 masses
phantom dimensions
50 % glandular
4.2 cm thick
dose less than 3mGy per image (+grid)
localization/accuracy (for stereo bx)
Daily before exams
Processor QC
Daily
Darkroom cleanliness
daily
viewbox conditions
weekly
phantom eval
weekly
repeat analysis
quarterly
compression test
semi-annually
darkroom fog
semi-annually
screen-film contrast
semi-annually
cancer per 1000 screeners
3-8
diagnostic mammo = mass
US = pissed off looking cancer
next step
stage. scan rest of boob for multi-focal masses AND axilla for nodes