Monitoring the Neuromuscular Junction Flashcards
What is the NMJ (Neuromuscular Junction)?
A synapse at which an electrical impulse traveling down a motor nerve, releases chemical transmitter which causes the muscle to contract
Steps to contraction:
- Impulse arrives at the motor end plate causing release of Ach from vesicles in axon terminus (Ca++ causes Ach to leave) 2. Ach diffuses across synapse to bind to receptors on the sarcolemma 3. Binding of Ach to receptors causes Na+ to enter cell and cause depolarization 4. If threshold depolarization occurs a new impulse moves along sarcolemma 5.Sarcolemma repolarizes (K+ leaves cell, Na/K pump restores polarity) 6. Ach broken down by Ach-E
Criteria for Nerve Selection:
-It must have a motor element -It must be close to skin -Contraction in the muscle or muscle group which the nerve supplies must be visible or accessible to evoked response monitoring
Criteria of nerve stimulator:
-Battery powered device that delivers depolarizing current via electrodes -Impulse < 0.5 msec and 0.1 sec in duration to elicit nerve firing at a readily attainable current (Pulse width > 0.5 msec extends beyond the refractory period of the nerve resulting in repetitive firing)
Pulse Width:
The duration of the individual impulse delivered by the nerve stimulator
Time to monitor:
-Time to intubation -Degree of relaxation -Time to reverse -Time for extubation -Residual curarization
Indications of Monitoring:
-Long interventions -Changed pharmacokinetics/dynamics -No moving/straining allowed -No reversal preferred -Disturbed electrolyte balance -Expected drug interactions
Techniques:
-Peripheral nerve stimulation (PNS) -Mechanomyograph (MMG) -Electromyograph (EMG) -Acceleromyography (AMG)
Peripheral Nerve Stimullus
-Visual and/or tactile -Muscle monitored should be in sight -Lacks accuracy and reliability -Acceptable TOF ratio of >70% for extubation (only 10% correct observations) -Double burst stimulus (DBS):Only 40% of anesthesiologists are able to recognize a fade
Mechno-myograph
-Isometric measurement of force of contraction with a force displacements transducer -Simple, accurate and reliable -Sensitive to external physical influences and limb has to be fixed in one position -Used for studies
Electromyography
-Measures evoked compound muscle action potential -Correct positioning of electrodes very important -Extensive and sensitive equipment -Diathermy interference -Seems to underestimate block during recovery -For scientific use
Acceleromyography
-Newton’s second law (F = M X a) -Transducer is easily placed but must move freely for reliable measurement
Stimulating patterns: Single Twitch (ST)
-Reflects events at post junctional membrane -Single supra maximal electrical stimuli applied to peripheral motor nerve -Frequency every second (1 Hz) or every 10 seconds (0.1 Hz) -Used for monitoring onset of block -Same response to both groups of NMBs -Response influenced by position of muscle, muscle temp -Calibration required before relaxation -Not for day to day clinical practice
Stimulating patterns: Train of Four (TOF)
-Reflects events at pre-synaptic membrane -Used successfully for onset, maintenance and recovery of block -4 supra maximal stimuli q 0.5 seconds (2 Hz) May be repeated q 12-15 seconds -Advantage: relative ration of 4th to 1st response remains the same despite changes in absolute responses
Non-Depolarizing Block Response: