Module 8: Other Photometric Methods Flashcards

1
Q

Flame emission photometry

A

certain elements (alkali metals) can emit characteristic wavelengths of light when headed in a flame
Na, K, Li most commonly evaluated
Sample introduced into a flame, atoms of elements are excited by heat energy (move into higher energy orbitals)
standards of known concentration are measured to make cal curve

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

when excited electrons return to the ground state,

A

they emit light energy at specific wavelengths characteristic for that element

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Flame emission photometry reaction

A

A+ + e- = Ao
Ao + heat = A*
A* = Ao + hv

A* = excited atom
Ao = atom in ground state
hv = photon of light energy
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

light vs atoms vs concentration in flame photometry

A

amount of light = number of atoms = concentration of element

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Na, K, Li wavelengths for flame photometry

A

Na measured at 589nm
K measured at 766nm
Li measured at 670nm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Components of flame photometer

A

Atomizer/burner assembly: acts as cuvet or sample holder; aspirates constant amount of solution into the flame as fine droplets (nebulization)
Wavelength selector: interference filter with narrow bandpass. Isolate wavelength characteristic of the element to be measured
Photodetector: photomultiplier tubes
Signal processor: amplify signal
Display

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Factors affecting the flame in flame photometry

A

1) type of fuel and oxidant (usually propane and air are used, flame is 1925degC, must be pure propane)
2) Fuel to oxidant ratio: determined by pressure regulation, required to produce constant thermal output

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Flame photometry interference: Spectral

A

other substances may emit light as well (controlled by using low flame temp and narrow bandpass filter)
Self-absorption when energy emitted is absorbed by other atoms of the same element (controlled by diluting samples to be measured)
Mutual excitation when transfer of energy from an atom of one element to an atom of another (controlled by using an internal standard that acts as a radiation buffer)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Flame photometry interference: ionization

A

if atoms are provided with sufficient energy they will ionize and emit different wavelengths of light (fewer atoms in the flame emitting the measured wavelength)
controlled by keeping flame temp low

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Flame photometry interference: physical interferences

A

Surface tension reduced by addition of wetting agent

Viscosity reduced by dilution of the sample (usually 1/100 or 1/200)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Internal standard

A

a solution with a known concentration of a substance that is structurally similar to the analytes of interest that is added to all samples
Used to account for variations in the system
Li or Cesium often used
Added to all samples and any variations in the system will affect the signal reading of both analytes and the internal standard

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Variation that an internal standard can account for

A

fluctuations in air/gas pressure causing changes in flame temp and stability
changes in aspiration rate affecting size and # of droplets
Mutual excitation of K atoms by Na atoms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

desirable characteristics of an internal standard element

A

high emission intensity
normally absent from biological fluids
emission at a wavelength sufficiently removed from Na and K to permit spectral isolation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Types of scattering

A

Rayleigh scatter
Raleigh-Debye scatter
Mie scatter (aka Tyndall or Tyndall-Mie)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Rayleigh scatter

A

particle size (d) is smaller than wavelength of incident light (d

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Raleigh-Debye scatter

A
particle size (d) is approximately equal to wavelength of incident light (d=λ)
as particle size increases, the scatter loses its symmetry and shows an increase in forward scattering
17
Q

Mie scatter (AKA Tyndall or Tyndall-Mie)

A
particle size (d) is larger than wavelength of incident light (d>λ)
large particles produce a predominance of scattered light in a narrow angular region in the forward direction
18
Q

What is used to produce consistent scatter patterns

A

a monochromator used to select specific wavelength of light

19
Q

2 photometric methods to measure scattered light

A

turbidimetry

nephelometry

20
Q

Turbidimetry

A

measurement of the decrease in intensity of the incident light beam that is caused by scattering
Inverse relationship between the light measured and the concentration of the sample
Same setup as photometer, with photodetector directly in line with the path of the light beam

21
Q

Nephelometry

A

measurement of the amount of light scattered
Direct relationship between the light measured and the concentration
Photodetector is NOT in the direct path of the light beam (instead placed at an angle to the light path, typically 90Degrees)
Multiple photodetectors at different angles may be used

22
Q

Sources of error for light scattering methods

A

Variations in particle size
Matrix effects (other substances in the solution that scatter light)
Debris within the system (dust)
Fluorescence (fluorescing compounds can be excited if short wavelength is used - false increase in light measured)
Settling of particles (mix prior to reading)

23
Q

Fluorescence

A

once the molecule is in excited state, radiationless vibrational deactivation (RVD) occurs until it is in its lowest excited state (some energy is lost)
As electron drop to ground state, light energy is emitted
Because some has already been lost to RVD, light energy emitted is less than and of a longer wavelength than the excitation energy
**Compounds that fluoresce absorb light of one wavelength and emit light of a longer wavelength

24
Q

Fluorescence excitation wavelength

A

in the UV range

25
Q

Fluorescence emission wavelength

A

in the visible range

26
Q

Fluorescence: Stokes shift

A

difference between the max wavelength of excitation light and max wavelength of emitted fluorescent light
Constant measure of the energy lost in the excited state due to RVD

27
Q

Clinical uses of fluorescence

A

often used as tags or cell markers
Hematology analyzer/flow cytometry to identify blood cells
Micro to measure bacterial growth
Immunoassay analyzer that use antibodies to detect specific drugs/substances

28
Q

Advantages to fluorometry

A

1000-10 000x more sensitive than photometry

More specific because only the wavelength emitted by the compound is measured

29
Q

Fluorometer

A

Fluorometer: use glass or interference filters
Spectrofluorometer: used prisms/diffraction gratings

measurement is made at 90deg to incident light (no interference from light transmitted by sample)
2 monochromatic devices: 1 to isolate the excitation wavelength (primary), one to isolate the emission wavelength (secondary)

30
Q

Components of fluorometers

A

Light source: 300-500nm, very high energy (xenon, quartz halogen, mercury), lasers may be used
Wavelength selector: filters, prisms, diffraction grating
Sample Holder: cuvet made of silica or quartz
Photodetector: photomultiplier tube (wide range of spectral responses, rapid response time, sensitive)

31
Q

Limitations of Fluorescence measurements

A

Inner filter effect: as concentration increases, relationship becomes nonlinear. Caused by loss of excitation intensity across the cuvet path length
Quenching: self absorption
Light scattering: scattered light reaching emission photodetector causes background interference (controlled with narrow bandpass filters)
Solvent effects: impurities in the solvent that may have fluorescent properties
Sample matrix effects: serum and urine may contain compounds that can fluoresce
Temp effects: fluorescence intensity decreases with increasing temp

32
Q

Reflectance photometry

A

incident beam of light is directed at test sample and reflected light is measured.
Amount of light reflected depends on concentration of analyte
Amount of light reflected back is inversely related to concentration

33
Q

Dry chemical systems

A

disposable dry reagent slides that also function as the cuvet
Spreading Layer: sample applied
Reagent/indicator layer: contain reagent required for the reaction. Analyte reacts with reagent to produce color change
Support layer: clear plastic support that allows incident light to pass through the indicator layer
Incident light is directred from underneath
Light strikes underside of spreading layer and is reflected back (reflected light measured by photodetector)

34
Q

Standardizing a dry chemical reflectance photometry system

A

0 concentration standard = white surface (100% reflectance)

high concentration standard = black surface (0 reflectance)