Module 3: The pacing system Flashcards

1
Q

Pacemaker?

A
  • small implantable device that helps control abnormally slow heart rhythms.
  • mimics the patient’s natural heart rhythm as much as possible.
  • two main functions: sensing the heart and pacing the heart.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Pacing system?

A

• Consists of an implantable pulse generator (IPG) = battery, circuitry, and connectors.
• One or more leads, which are wires that conduct, or carry, the impulses to the heart.
o A cathode = electrode with a negative charge that delivers the electrical impulse to the myocardium.
o An anode = electrode with a positive charge to which the impulse returns after stimulating the heart.

• Body tissue and fluids + IPG + leads + cathode + anode = conduction of electricity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Purpose of pacing system

A
  • Form a conduction pathway to pace the heart when necessary

* Form a sensing circuit to sense intrinsic cardiac activity, withholding pacing when appropriate.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Purpose of a telemetry coil

A
  • establish a connection with a Medtronic programmer
  • Once connected
    o Sends programming instructions down to the device.
    o Sends diagnostic data up to the programmer
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Purpose of sensors

A
  • Measure the patient’s movement and level of exertion. (Info is used to change the heart rate so that it is appropriate for the patient’s current activity level.)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Purpose of Leads, or insulated conductor wires

A
  • Deliver electrical impulses to the heart.
  • Sense the electrical signals of the patient’s own heart beat. One end of the lead connects to the pulse generator, and the other is in contact with cardiac tissue
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Pacemaker component: Circuit board

A
  • Contains microprocessors that allow both sending programmed instructions to, and receiving and analyzing data from, the heart.
    Circuits permit:
  • Programming of pacing therapies and diagnostic functions by establishing telemetry or communication with the programmer
  • the integrated circuit’s ability to sense the heart’s electrical activity
  • Collecting diagnostic data of both pacing system operation and the patient’s intrinsic heart activity
  • Components in the circuit board
    o Resistors = control the flow of electricity
    o The defibrillation protection unit = allows the pacemaker to safely accept high voltage external defibrillation without destroying the other components
    o The output capacitors and “voltage doubler,” = allow the pacemaker to deliver varying voltages
    o Timers = keep device operation and function consistent within the pacemaker
    o The reed switch = allows mode and rate changes when a magnet is placed near the device
    o The telemetry antenna = enables communication between the programmer and the pacemaker
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Pacemaker component: Battery

A
  • Provides energy for sending electrical impulses to the heart. Lithium-iodine is the most commonly used power source for today’s pacemakers (an initial open circuit voltage of 2.8 volts).
  • ICDs use Silver-Vanadium-Oxide cells with a higher voltage
  • Functions
    o Provides the power source for the device’s multiple functions.
    o Provide the energy needed to send electrical impulses to the heart.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Pacemaker component: Header block

A
  • Connects the lead to the pacemaker > the device can sense what is happening in the patient’s heart and pace when necessary.
  • In a dual-chamber pacemaker, the atrial and ventricular connectors each have their own housings in the connector block.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Leads?

A
  • Insulated wires that deliver electrical impulses from the pulse generator to the heart, and sense cardiac depolarization.
  • Lead is subject to mechanical torque, flexing, bending, and the body’s natural defenses, including bio-chemicals produced in response to inflammation.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Fixation Mechanism: Active leads

A

Better choice to prevent lead dislodgment in patients:

  • With smooth-walled hearts
  • Whose hearts lack trabeculation, such as dilated hearts
  • Who have had a previous CABG procedure
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Where is active lead tip screwed to? Adv?

A

Lead tip is screwed to the endocardial wall.

  • Fixation mechanism secured into heart muscle
  • Creates a solid connection between the lead tip and the heart > mitigates the risk of lead dislodgements
  • positioned almost anywhere in the heart.
  • Makes active fixation leads easier to remove (when necessary) because the lead tip can easily be unscrewed back up into the helix of the lead at any time.
  • Steroid eluting > reduce tissue inflammation and heal the cardiac tissue, negating the consequences of inflamed cardiac tissue caused by the sharp screw.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Where is passive lead tip screwed to?

A

Implanted where there is a atrial appendage, or where there is trabeculae in the ventrical. Cannot be implanted on areas where the heart wall is smooth.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Implantation of a passive lead tip

A

During implantation, the lead tip is lodged between the web-like pectinate of the atrium or trebeculae of the ventricle.

Once the lead tip is in place, the tines, made of soft silicone, get “hung up” in the pectinate or trebeculae, helping to keep the lead in place. After roughly 4 to 8 weeks, a fibrotic meshwork of the heart tissue encapsulates the tined lead tip. At this point, the lead is extremely secure and unlikely to dislodge. However, if the lead needs to be removed after this time, it is very difficult to do so, and the patient needs a very specialized procedure to get the lead removed.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Adv of passive leads

A
  • less tissue damage to the heart
  • Becomes lodged in trabeculae
  • Extremely secure once fibrotic tissue locks the lead in place
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Epicardial leads

A

implants represent less than 5% of leads implanted
primarily in pediatric cases
applied directly to the surface of the heart

17
Q

Insulation material: Silcone leads

A
  • Reliable
  • Relatively fragile, must be thicker
  • Higher coefficient of friction
  • Medtronic’s silicure has improved mechanical strength
  • Inert
  • Biocompatible
  • Biostable
  • Repairable with medical adhesive
18
Q

Insulation material: Polyurethane leads

A
  • Slightly more durable
  • Allows thinner applications when compared to silicone
  • more lubricious due to a lower friction coefficient. This characteristic can make it easier to implant.
  • Early types of urethane were subject to chemical reactions
  • Higher tear strength
  • slightly smaller in diameter.
19
Q

Unipolar lead

A

consist of a single conductor coil extending to the tip electrode, or cathode

20
Q

Bipolar lead

A

two coils—one to the tip cathode, the other to the ring-anode. The coils are separated by an insulating layer.

21
Q

Unipolar Pacing

A
  • Travels down the lead wire to stimulate the heart at the tip electrode, which is also referred to as the cathode;
  • Returns to the metal casing of the impulse generator, or the anode, by way of body fluids.
  • producing larger spikes on the ECG. Since the electrodes are located at a relatively large distance from each other, detected surface ECG signals are usually larger.
  • Another characteristic is producing extracardiac stimulation in the form of muscle stimulation or pacemaker pocket stimulation, which is when pacemaker current flows between the tip and the can.
22
Q

Bipolar Pacing

A
  • the impulse travels down the inner wire of the lead to stimulate the heart at the tip electrode, or the cathode.
  • The impulse then travels to the ring electrode, or the anode, which is located several millimeters above the lead tip.
  • From the anode, the impulse travels up the outer wire of the lead to the pulse generator and completes the circuit.
  • There is less likelihood of extracardiac stimulation, such as pocket stimulation or muscle stimulation. This is because current returns to the ring electrode within the heart, rather than to the can as in unipolar pacing.
  • inability to see pacing spikes on an ECG. This is because the current pathway is so short between the electrodes.
23
Q

Bipolar Sensing

A
  • A bipolar system is less susceptible to oversensing non-cardiac signals, such as myopotentials, far-field intercardiac signals, noise, and EMI. This is because the “antenna,” or the distance between the two electrodes, is small and insulated within the chamber of the heart that is being sensed.
24
Q

Unipolar Sensing

A
  • susceptibility to oversensing. This is due to the big distance between electrodes. Extracardiac signals, such as myopotential signals or Electromagnetic Interference (EMI) outside the body, could potentially be sensed. Older IPGs were more susceptible to sensing extracardiac signals, but today’s modern IPGs have better sense filters to avoid this problem.
25
Q

Biopolar system: lead characteristics

A
  • Larger diameters. Over the years, however, bipolar sensing systems have grown smaller due to coaxial conductor designs, advanced thinner insulations, and smaller conductor coils and cables. Note also that the second conductor coil and insulation between the conductor coils make bipolar leads stiffer. A softer tip-to-ring spacer is used in many lead designs to make the distal tip more flexible.
  • built-in safety mechanism.” If the lead developed a problem, such as with the outer insulation or outer wire, the lead could be programmed to unipolar and still function for the patient. Thus, bipolar leads provide more alternatives when the lead experiences problems.
26
Q

Unipolar system: lead characteristics

A
  • smaller diameters > one conductor coil, so the insulator between the two conductor coils is eliminated.
  • less rigidity > only one conductor coil and one layer of insulation.
27
Q

Adv/Disadv of unipolar system

A

Advantages of the unipolar system include:
• The sensing of smaller cardiac signals that may not be detected in bipolar;
• The larger pacing spikes on the ECG; and
• The smaller diameter of the leads when compared to bipolar leads.
A disadvantage of the unipolar system
- may over-sense myopotentials
- “far-field” intracardiac signals, noise, and EMI.

28
Q

2 key clinical goals?

A
  • To deliver sufficient energy (including a safety margin) to consistently cause the heart to contract when necessary; and
  • Maximizing battery performance so the device lasts as long as possible.
29
Q

Pacing systems focus on 2 key relationships?

A
  1. Voltage and current drain; and

2. Impedance and current drain

30
Q

What is current drain?

A

The total amount of electrical capacity used in a pacing system over a given period of time.

31
Q

Current? Voltage? Impedance?

A

Current: Rate at which electricity travels through a circuit
Impedance: Resistance to the flow of electricity
Voltage: The force that pushes on electrons causing them to move through a pacing circuit