What resolves aliasing
Recall aliasing is when the Doppler shift is no more than the nyquist limit which is 1/2 PRF
decreasing the depth of the sample volume (answer) because it increases PRF (increasing nyquist limit)
Do you increase or decrease operator frequency to resolve aliasing
Decrease operator frequency (which would decrease Doppler shift)
You want to decrease Doppler shift OR
increase PRF
Because a Doppler shift if anything over 1/2 PRF will alias
The greatest portion of circulating blood is located in the
Venous system
True or false
Size of the sample volume is determined by ensemble length
number of pulses per scan line does not change window size
Spatial pulse length
Number of cycles in a pulse
The greater the impedance difference the greater the
Reflection
NOT refraction
Refraction requires two things to happen
Different propagation speeds Oblique incidence (not perpendicular)