What is the preferred light source for MIS surgery?
Xenon, as it emits a high intensity light that closely reproduces the color of natural light (as compared to halogen)
What level of mechanical insufflation in the thoracic cavity causes significant cardiorespiratory depression?
3 mmHg
What intra-abdominal pressure has been shown to result in acceptable levels of cardiorespiratory pressure and adequate working space in dogs?
15 mmHg - but levels of 8-10 mmHg are often adequate
What level of intra-abdominal pressure is tolerated in cats?
Study showed no significant cardiopulmonary compromise with intraabdominal pressures up to 8 mmHg, with little increase in working volume when pressures increased above this level
What is the most common gas used for insufflation?
CO2 - cheap, colorless, unlikely to combust, highly soluble so unlikely to form gas emboli, rapidly excreted from the body
What size vessels is the Harmonic ace tip vessel sealing device indicated for?
Up to 5mm
What are the available lengths and staple leg lengths of endoGIA stapling devices?
Lengths: 30, 45, 60mm
Leg lengths: 2.0, 2.5, 3.5, 4.8 (4.8 requires passage down a 15mm cannula)
What limits perfusion to the collapsed lung in one lung ventilation?
Hypoxic pulmonary vasoconstriction